J Cataract Refract Surg. 2011 Feb;37(2):324-7. doi: 10.1016/j.jcrs.2010.08.033.
To assess the safety to the corneal endothelium of ultraviolet (UV) light treatment to a light-adjustable intraocular lens (IOL).
Codet Vision Institute, Tijuana, Mexico.
Cohort study.
The study evaluated consecutive patients scheduled for phacoemulsification cataract surgery with implantation of a light-adjustable IOL. Endothelial cell density (ECD), variation in the size of endothelial cells, and the percentage of hexagonal cells were determined centrally with a specular microscope preoperatively, 1 week postoperatively (before UV treatment), and 6 months postoperatively. Two adjustments and 2 lock-in procedures were performed with a light-delivery system between 2 weeks and 4 weeks postoperatively.
The study enrolled 10 patients. Specular microscopy showed a mean ECD of 2344 cells/mm(2) preoperatively. The mean overall central ECD was 2047 cells/mm(2) 1 week postoperatively and 2130 cells/mm(2) at 6 months, representing an endothelial cell loss of 12.6% and 9.1%, respectively, from preoperative values. There was no statistically significant difference in the coefficient of variation or percentage of hexagonal cells at either postoperative evaluation.
The endothelial cell loss at 6 months correlated well with previous reports of endothelial damage after phacoemulsification with IOL implantation. The light-adjustable IOL irradiation protocol for adjustment and lock-in procedures did not seem to add to the endothelial damage caused by the cataract surgery, indicating the protocol is safe to the corneal endothelium.
评估可调光型人工晶状体(IOL)紫外线(UV)光治疗对角膜内皮细胞的安全性。
墨西哥提华纳 Codet Vision 研究所。
队列研究。
该研究评估了连续接受超声乳化白内障手术并植入可调光 IOL 的患者。使用共焦显微镜在术前、术后 1 周(UV 治疗前)和术后 6 个月,对中央角膜内皮细胞密度(ECD)、内皮细胞大小变化和六边形细胞百分比进行测量。术后 2 至 4 周内,使用光传输系统进行 2 次调整和 2 次锁定程序。
该研究共纳入 10 例患者。术前共焦显微镜显示平均 ECD 为 2344 个细胞/mm²。术后 1 周平均总中央 ECD 为 2047 个细胞/mm²,术后 6 个月为 2130 个细胞/mm²,分别较术前值损失 12.6%和 9.1%。术后评估时,变异系数或六边形细胞百分比均无统计学差异。
术后 6 个月的内皮细胞丢失与白内障超声乳化联合 IOL 植入术后内皮损伤的先前报告相符。调整和锁定程序的可调光 IOL 照射方案似乎不会增加白内障手术引起的内皮损伤,表明该方案对角膜内皮安全。