University of Wisconsin-Madison School of Medicine and Public Health, Department of Surgery, Division of Otolaryngology-Head and Neck Surgery, Madison, Wisconsin 53706, USA.
Laryngoscope. 2011 Apr;121(4):793-800. doi: 10.1002/lary.21431. Epub 2011 Feb 8.
We present the adjustable balloon implant (ABI), a novel implant to be used in type I thyroplasty for the treatment of vocal fold paralysis. The ABI offers the same medialization provided by other implants, but can easily be catered to individual patient anatomy as well as modified postoperatively without the need for a revision thyroplasty.
Repeated measures with each larynx serving as its own control.
Medialization thyroplasty (MT) with the ABI was performed on five excised canine larynges. Mucosal wave, aerodynamic, and acoustic parameters were measured for three conditions: normal; right vocal fold paralysis; and paralysis with the ABI.
Insertion of the ABI resulted in significant decreases in both phonation threshold pressure and phonation threshold flow. Perturbation parameters of percent jitter and percent shimmer were also significantly decreased and restored to normal levels. Signal-to-noise ratio was significantly increased to the normal level as well. The mucosal wave was preserved after implant insertion.
This preliminary experiment showing significant improvements in aerodynamic and acoustic parameters demonstrates the potential of the ABI as a thyroplasty implant. Effective medialization and preservation of the mucosal wave combined with postoperative adjustability makes it a potentially valuable clinical device.
我们介绍了一种新型的可调节球囊植入物(ABI),可用于治疗声带麻痹的 I 型甲状软骨成形术。ABI 提供了与其他植入物相同的内移效果,但可以根据患者的个体解剖结构轻松调整,并在术后无需进行翻修甲状软骨成形术的情况下进行修改。
每个喉作为自身对照的重复测量。
在五个切除的犬喉中进行 ABI 的内侧化甲状软骨成形术(MT)。测量了三种情况的黏膜波、气动和声学参数:正常;右侧声带麻痹;以及 ABI 伴声带麻痹。
ABI 的插入导致发声阈值压力和发声阈值流量显著降低。抖动百分比和闪烁百分比的扰动参数也显著降低,并恢复到正常水平。信噪比也显著增加到正常水平。黏膜波在植入物插入后得以保留。
这项初步实验显示气动和声学参数的显著改善,证明了 ABI 作为甲状软骨成形术植入物的潜力。有效的内移和黏膜波的保留结合术后的可调节性,使其成为一种具有潜在价值的临床设备。