Division of Laboratory Medicine, Boston Medical Center, 88 East Newton Street, H3600, Boston, MA 02118, USA.
J Clin Microbiol. 2011 Apr;49(4):1624-7. doi: 10.1128/JCM.01958-10. Epub 2011 Feb 9.
Antibiotic neutralization in blood culture media from two automated systems was evaluated by measuring the recovery of organisms and times to detection in simulated cultures. Overall, BD Bactec Plus media (Bactec FX system) outperformed TREK 80 ml Redox media (VersaTREK system), although results suggest a relative rather than an absolute increased rate of recovery for the Bactec media.
通过测量模拟培养物中微生物的回收情况和检测时间,评估了两种自动化系统的血培养物培养基中的抗生素中和作用。总体而言,BD Bactec Plus 培养基(Bactec FX 系统)优于 TREK 80 ml Redox 培养基(VersaTREK 系统),尽管结果表明 Bactec 培养基的回收率相对提高,而非绝对提高。