Department of Human Oncology, School of Medicine and Public Health, University of Wisconsin-Madison, 600 Highland Avenue, Madison, WI, USA.
Expert Rev Anticancer Ther. 2011 Feb;11(2):309-20. doi: 10.1586/era.10.175.
As a modality for delivering rotational therapy, helical tomotherapy offers dosimetric advantages by combining a continuously rotating gantry with a binary multileaf collimator. Helical tomotherapy, embodied in the TomoTherapy(®) Hi-Art II(®) system, delivers intensity-modulated fan beams in a helical pattern using binary multileaf collimator leaves while the couch is translated through the gantry. Helical tomotherapy offers the possibility of treating a variety of cases--from simple to complex--with improved target conformality and sensitive structure sparing compared with 3D or conventional static field IMRT plans, thereby allowing biologically effective dose escalation. For precise irradiation and possible treatment adaptation, the fully integrated on-board image-guidance system provides online volumetric images of patient anatomy using 3.5-MV x-ray beams and the xenon computed tomography detector. Several review articles were published before the year 2007 but emphasized the technical aspects of helical tomotherapy. In this article, we review very recent papers and focus on the dosimetric and clinical aspects of helical tomotherapy.
作为一种提供旋转治疗的方式,螺旋断层放疗通过将连续旋转机架与二进制多叶准直器相结合,提供了剂量学优势。螺旋断层放疗体现在 TomoTherapy(®)Hi-Art II(®)系统中,通过使用二进制多叶准直器叶片在机架旋转的同时在螺旋模式下输送强度调制扇形束,同时治疗床在机架中平移。与 3D 或常规静态场调强放疗计划相比,螺旋断层放疗提供了治疗各种病例的可能性——从简单到复杂,具有更好的靶区适形性和敏感结构保护,从而允许生物有效剂量提升。为了实现精确照射和可能的治疗适应,完全集成的机载图像引导系统使用 3.5-MV X 射线束和氙计算断层摄影探测器提供患者解剖结构的在线容积图像。在 2007 年之前,已经发表了几篇综述文章,但强调了螺旋断层放疗的技术方面。在本文中,我们回顾了最近的论文,并重点介绍了螺旋断层放疗的剂量学和临床方面。