Suppr超能文献

采用三维各向同性分辨率快速自旋回波 MRI 评估臂丛神经:与 3.0T 常规 MRI 的比较。

Brachial plexus assessment with three-dimensional isotropic resolution fast spin echo MRI: comparison with conventional MRI at 3.0 T.

机构信息

Department of Radiology, National Institute for Cancer Research, Genoa, Italy.

出版信息

Br J Radiol. 2012 Jun;85(1014):e110-6. doi: 10.1259/bjr/28972953. Epub 2011 Feb 22.

Abstract

OBJECTIVE

The purpose of our study was to determine whether a three-dimensional (3D) isotropic resolution fast spin echo sequence (FSE-cube) has similar image quality and diagnostic performance to a routine MRI protocol for brachial plexus evaluation in volunteers and symptomatic patients at 3.0 T. Institutional review board approval and written informed consent were guaranteed.

METHODS

In this prospective study FSE-cube was added to the standard brachial plexus examination protocol in eight patients (mean age, 50.2 years) with brachial plexus pathologies and in six volunteers (mean age, 54 years). Nerve visibility, tissue contrast, edge sharpness, image blurring, motion artefact and acquisition time were calculated for FSE-cube sequences and for the standard protocol on a standardised five-point scale. The visibility of brachial plexus nerve and surrounding tissues at four levels (roots, interscalene area, costoclavicular space and axillary level) was assessed.

RESULTS

Image quality and nerve visibility did not significantly differ between FSE-cube and the standard protocol (p>0.05). Acquisition time was statistically and clinically significantly shorter with FSE-cube (p<0.05). Pathological findings were seen equally well with FSE-cube and the standard protocol.

CONCLUSION

3D FSE-cube provided similar image quality in a shorter acquisition time and enabled excellent visualisation of brachial plexus anatomy and pathology in any orientation, regardless of the original scanning plane.

摘要

目的

我们的研究目的是确定在 3.0T 下,三维各向同性分辨率快速自旋回波序列(FSE-cube)是否与常规 MRI 方案在志愿者和症状性患者的臂丛神经评估中具有相似的图像质量和诊断性能。本研究获得了机构审查委员会的批准和书面知情同意。

方法

在这项前瞻性研究中,我们在 8 例(平均年龄 50.2 岁)臂丛神经病变患者和 6 例志愿者(平均年龄 54 岁)中,将 FSE-cube 添加到标准臂丛神经检查方案中。我们使用标准化的五分制量表计算 FSE-cube 序列和标准方案的神经可见度、组织对比度、边缘锐度、图像模糊、运动伪影和采集时间。评估了四个水平(神经根、斜角肌间隙、肋锁间隙和腋窝水平)的臂丛神经和周围组织的可见度。

结果

FSE-cube 和标准方案之间的图像质量和神经可见度没有显著差异(p>0.05)。FSE-cube 的采集时间在统计学和临床意义上明显更短(p<0.05)。FSE-cube 和标准方案都可以同样清晰地显示病变发现。

结论

3D FSE-cube 在更短的采集时间内提供了相似的图像质量,并能够以任何方向极好地显示臂丛神经解剖结构和病理学,而无需考虑原始扫描平面。

相似文献

3
A robust 3D fast spin-echo technique for fast examination of the brachial plexus.
Skeletal Radiol. 2022 Sep;51(9):1865-1872. doi: 10.1007/s00256-022-04021-9. Epub 2022 Mar 26.
4
Ankle: isotropic MR imaging with 3D-FSE-cube--initial experience in healthy volunteers.
Radiology. 2008 Dec;249(3):1026-33. doi: 10.1148/radiol.2493080227.
6
Imaging of the wrist at 1.5 Tesla using isotropic three-dimensional fast spin echo cube.
J Magn Reson Imaging. 2011 Apr;33(4):908-15. doi: 10.1002/jmri.22494.
8
T2-weighted IDEAL fast spin echo imaging of the brachial plexus: comparison with STIR.
Acta Radiol. 2015 Oct;56(10):1242-7. doi: 10.1177/0284185114552292. Epub 2014 Oct 15.
10
MR imaging of the brachial plexus: comparison between 1.5-T and 3-T MR imaging: preliminary experience.
Skeletal Radiol. 2011 Jun;40(6):717-24. doi: 10.1007/s00256-010-1050-x. Epub 2010 Oct 23.

引用本文的文献

2
The guiding value of the cinematic volume rendering technique in the preoperative diagnosis of brachial plexus schwannoma.
Front Oncol. 2023 Dec 13;13:1278386. doi: 10.3389/fonc.2023.1278386. eCollection 2023.
4
An optimized 1.5 Tesla MRI protocol of the brachial plexus.
Neuroradiol J. 2024 Feb;37(1):43-53. doi: 10.1177/19714009231196475. Epub 2023 Aug 24.
5
A pentavalent approach for the evaluation of traumatic brachial plexopathy on MRI: correlation of macropattern and micropattern.
Br J Radiol. 2023 Jun 1;96(1146):20220913. doi: 10.1259/bjr.20220913. Epub 2023 Mar 1.
6
The Application of Contrast-Enhanced 3D-STIR-VISTA MR Imaging of the Brachial Plexus.
J Belg Soc Radiol. 2022 Sep 5;106(1):75. doi: 10.5334/jbsr.2803. eCollection 2022.
7
Single-direction diffusion-weighted imaging may be a simple complementary sequence for evaluating fetal corpus callosum.
Eur Radiol. 2022 Feb;32(2):1135-1143. doi: 10.1007/s00330-021-08176-2. Epub 2021 Jul 30.
9
Peripheral nerve imaging: Not only cross-sectional area.
World J Radiol. 2016 Aug 28;8(8):726-8. doi: 10.4329/wjr.v8.i8.726.

本文引用的文献

1
Evaluation of uterine anomalies: 3D FRFSE cube versus standard 2D FRFSE.
AJR Am J Roentgenol. 2009 Dec;193(6):W558-62. doi: 10.2214/AJR.09.2716.
3
Elbow nerves: MR findings in 60 asymptomatic subjects--normal anatomy, variants, and pitfalls.
Radiology. 2009 Jul;252(1):148-56. doi: 10.1148/radiol.2521081614. Epub 2009 May 18.
4
Turboprop IDEAL: a motion-resistant fat-water separation technique.
Magn Reson Med. 2009 Jan;61(1):188-95. doi: 10.1002/mrm.21825.
5
Ankle: isotropic MR imaging with 3D-FSE-cube--initial experience in healthy volunteers.
Radiology. 2008 Dec;249(3):1026-33. doi: 10.1148/radiol.2493080227.
6
Effects of refocusing flip angle modulation and view ordering in 3D fast spin echo.
Magn Reson Med. 2008 Sep;60(3):640-9. doi: 10.1002/mrm.21680.
9
Clinical applications of diffusion tensor tractography of the spinal cord.
Neuroradiology. 2008 Jan;50(1):25-9. doi: 10.1007/s00234-007-0309-y. Epub 2007 Oct 2.
10

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验