Department of Radiology, National Institute for Cancer Research, Genoa, Italy.
Br J Radiol. 2012 Jun;85(1014):e110-6. doi: 10.1259/bjr/28972953. Epub 2011 Feb 22.
The purpose of our study was to determine whether a three-dimensional (3D) isotropic resolution fast spin echo sequence (FSE-cube) has similar image quality and diagnostic performance to a routine MRI protocol for brachial plexus evaluation in volunteers and symptomatic patients at 3.0 T. Institutional review board approval and written informed consent were guaranteed.
In this prospective study FSE-cube was added to the standard brachial plexus examination protocol in eight patients (mean age, 50.2 years) with brachial plexus pathologies and in six volunteers (mean age, 54 years). Nerve visibility, tissue contrast, edge sharpness, image blurring, motion artefact and acquisition time were calculated for FSE-cube sequences and for the standard protocol on a standardised five-point scale. The visibility of brachial plexus nerve and surrounding tissues at four levels (roots, interscalene area, costoclavicular space and axillary level) was assessed.
Image quality and nerve visibility did not significantly differ between FSE-cube and the standard protocol (p>0.05). Acquisition time was statistically and clinically significantly shorter with FSE-cube (p<0.05). Pathological findings were seen equally well with FSE-cube and the standard protocol.
3D FSE-cube provided similar image quality in a shorter acquisition time and enabled excellent visualisation of brachial plexus anatomy and pathology in any orientation, regardless of the original scanning plane.
我们的研究目的是确定在 3.0T 下,三维各向同性分辨率快速自旋回波序列(FSE-cube)是否与常规 MRI 方案在志愿者和症状性患者的臂丛神经评估中具有相似的图像质量和诊断性能。本研究获得了机构审查委员会的批准和书面知情同意。
在这项前瞻性研究中,我们在 8 例(平均年龄 50.2 岁)臂丛神经病变患者和 6 例志愿者(平均年龄 54 岁)中,将 FSE-cube 添加到标准臂丛神经检查方案中。我们使用标准化的五分制量表计算 FSE-cube 序列和标准方案的神经可见度、组织对比度、边缘锐度、图像模糊、运动伪影和采集时间。评估了四个水平(神经根、斜角肌间隙、肋锁间隙和腋窝水平)的臂丛神经和周围组织的可见度。
FSE-cube 和标准方案之间的图像质量和神经可见度没有显著差异(p>0.05)。FSE-cube 的采集时间在统计学和临床意义上明显更短(p<0.05)。FSE-cube 和标准方案都可以同样清晰地显示病变发现。
3D FSE-cube 在更短的采集时间内提供了相似的图像质量,并能够以任何方向极好地显示臂丛神经解剖结构和病理学,而无需考虑原始扫描平面。