Suppr超能文献

20兆赫多普勒导管冠状动脉测速法测量冠状动脉血流速度的准确性。

Accuracy of 20-MHz Doppler catheter coronary artery velocimetry for measurement of coronary blood flow velocity.

作者信息

Tadaoka S, Kagiyama M, Hiramatsu O, Ogasawara Y, Tsujioka K, Wada Y, Sawayama T, Kajiya F

机构信息

Department of Cardiology, Kawasaki Medical School, Kurashiki, Japan.

出版信息

Cathet Cardiovasc Diagn. 1990 Mar;19(3):205-13. doi: 10.1002/ccd.1810190313.

Abstract

The accuracy of 20-MHz Doppler catheter coronary artery velocimetry (Millar Instruments) was evaluated by model experiments using a turntable and a flow tube. In the turntable experiments, the flow away from the catheter tip (away flow) was significantly lower than that toward the catheter (toward flow), but the measured velocities for both away and toward flows were always lower than the known velocities. Nevertheless, the blood velocities measured at a sampling point 4 mm from the catheter tip showed excellent correlation with the known velocities irrespective of flow direction. The shape of the FFT spectra of the Doppler shift frequency from blood was broadened, and the frequency corresponding to the known velocity was always close to the maximum Doppler shift frequency. We concluded that zero-cross detection of blood velocity by the 20-MHz pulsed Doppler catheter provides a reliable measure for evaluating relative velocity changes, although it underestimates the blood velocity.

摘要

使用转盘和流管通过模型实验评估了20兆赫多普勒导管冠状动脉测速法(米勒仪器公司)的准确性。在转盘实验中,远离导管尖端的血流(远心血流)明显低于朝向导管的血流(近心血流),但远心血流和近心血流的测量速度始终低于已知速度。然而,在距导管尖端4毫米处的采样点测量的血流速度与已知速度显示出极好的相关性,与血流方向无关。血液多普勒频移频率的FFT频谱形状变宽,与已知速度对应的频率始终接近最大多普勒频移频率。我们得出结论,尽管20兆赫脉冲多普勒导管对血流速度的过零检测低估了血流速度,但它为评估相对速度变化提供了一种可靠的测量方法。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验