Graham S P, Cohen M D, Hodgson J M
Division of Cardiology, Medical College of Virginia, Richmond.
Cathet Cardiovasc Diagn. 1990 Mar;19(3):214-21. doi: 10.1002/ccd.1810190314.
This prospective study investigated two different methods used clinically to estimate coronary flow reserve during cardiac catheterization. First, the relationship between simultaneous digital angiography and intracoronary Doppler velocity measurements was determined in 61 patients. The correlation coefficient for 233 simultaneous pairs of papaverine-induced flow reserve was 0.70 (Doppler = 0.41 digital + 0.93, P less than 0.0001). Repeated basal flow determinations showed little variability (Doppler: 2 +/- 5% (mean +/- 95% confidence intervals); digital: 3 +/- 10%). Repeated hyperemic flow estimates by the Doppler technique were more reproducible (10 +/- 2%) than those by the digital method (26 +/- 6%). Second, estimates of coronary flow reserve by side-mounted and end-mounted Doppler catheters were compared in vitro and in patients. In vitro measurement of blood velocity was linear for both catheter designs and was highly correlated with volume flow determined by electromagnetic flow meter (r = 0.99 side-mounted; r = 0.96 end-mounted). In patients (n = 20), average coronary flow reserve for the side-mounted Doppler crystal was 2.64 +/- 0.15 and for the end-mounted Doppler crystal was 2.40 +/- 0.14 (P less than 0.02). When flow reserve was determined twice for each catheter, there was greater variability in repeated measurements using the side-mounted Doppler crystal (21 +/- 10%, n = 14) than using the end-mounted Doppler crystal (12 +/- 4%, n = 32).(ABSTRACT TRUNCATED AT 250 WORDS)
这项前瞻性研究调查了临床上用于在心脏导管插入术期间估计冠状动脉血流储备的两种不同方法。首先,在61例患者中确定了同步数字血管造影与冠状动脉内多普勒流速测量之间的关系。233对同时进行的罂粟碱诱导的血流储备的相关系数为0.70(多普勒=0.41×数字+0.93,P<0.0001)。重复进行的基础血流测定显示变异性很小(多普勒:2±5%(平均值±95%置信区间);数字:3±10%)。通过多普勒技术重复进行的充血血流估计比数字方法更具可重复性(10±2%)(26±6%)。其次,在体外和患者中比较了侧装式和端装式多普勒导管对冠状动脉血流储备的估计。两种导管设计的体外血流速度测量均呈线性,且与电磁流量计测定的体积流量高度相关(侧装式r=;端装式r=)。在患者(n=20)中,侧装式多普勒探头的平均冠状动脉血流储备为2.64±0.15,端装式多普勒探头为2.40±0.14(P<0.02)。当对每个导管进行两次血流储备测定时,使用侧装式多普勒探头重复测量的变异性(21±10%,n=14)大于使用端装式多普勒探头(12±4%,n=32)。(摘要截断于250字) (注:原文中侧装式和端装式r值未给出具体数字)