用于测量血流速度的侧向安装与端部安装冠状动脉内多普勒血流探头之间的显著差异。

Significant differences between side-mounted and end-mounted intracoronary Doppler flow probes for the measurement of blood flow velocity.

作者信息

Holdright D R, Clarke D, Ford D W, Fox K M, Poole-Wilson P A, Collins P

机构信息

National Heart & Lung Institute, London, England.

出版信息

Angiology. 1995 Jul;46(7):583-90. doi: 10.1177/000331979504600705.

Abstract

Use of the intracoronary Doppler flow probe is an established method for the assessment of coronary blood flow velocity. The aim of this study was to perform an in vitro comparison of two commonly used Doppler probes, which differ in the location of the piezoelectric crystal (end-mounted vs side-mounted). Blood flow velocity was measured over a wide range of flow rates in a flow simulator using heparinized whole blood. Measurements were made with both Doppler probes assessed in two positions (supported and unsupported) within the tubing. The results were compared with estimated true velocities. Further measurements were made with six side-mounted probes, correcting for the assumed crystal mounting angle and for the angle calculated from magnified images of the individual crystals. Mean velocities for end- and side-mounted probes correlated highly with predicted velocities (all r > or = 0.99), but the side-mounted probes significantly overestimated velocity by > 100%. Estimation of the true crystal mounting angle of the side-mounted probe revealed considerable variability (range 30-42 degrees) and was lower than the recommended angle correction factor of 60 degrees. Velocities corrected for the individual crystal mounting angles agreed more closely with predicted mean velocities. Although both probes are adequate for the assessment of relative changes in flow, the side-mounted probe considerably overestimates mean velocity, which is partly explained by the variable mounting angle of the crystal. The demonstrated limitations of the side-mounted Doppler flow probe in vitro should be considered in undertaking measurement of intracoronary blood flow velocity.

摘要

冠状动脉内多普勒血流探头的使用是评估冠状动脉血流速度的一种既定方法。本研究的目的是对两种常用的多普勒探头进行体外比较,这两种探头的压电晶体位置不同(端部安装式与侧面安装式)。在血流模拟器中使用肝素化全血在很宽的流速范围内测量血流速度。在管道内的两个位置(支撑和无支撑)对两种多普勒探头进行测量。将结果与估计的真实速度进行比较。对六个侧面安装式探头进行了进一步测量,校正了假定的晶体安装角度以及根据单个晶体的放大图像计算出的角度。端部安装式和侧面安装式探头的平均速度与预测速度高度相关(所有r≥0.99),但侧面安装式探头显著高估速度超过100%。对侧面安装式探头真实晶体安装角度的估计显示出相当大的变异性(范围为30 - 42度),且低于推荐的60度角度校正系数。针对单个晶体安装角度校正后的速度与预测平均速度更为接近。尽管两种探头都足以评估血流的相对变化,但侧面安装式探头显著高估了平均速度,这部分是由晶体可变的安装角度所解释的。在进行冠状动脉血流速度测量时,应考虑侧面安装式多普勒血流探头在体外所显示的局限性。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验