用激光多普勒流量计比较红光和红外线记录人前牙牙髓血流。

A comparison between red and infrared light for recording pulpal blood flow from human anterior teeth with a laser Doppler flow meter.

机构信息

Department of Physiology and Biochemistry, Faculty of Dentistry, Mahidol University, Yothi Street, Bangkok 10400, Thailand.

出版信息

Arch Oral Biol. 2011 Jun;56(6):614-8. doi: 10.1016/j.archoralbio.2011.02.010. Epub 2011 Mar 12.

Abstract

OBJECTIVE

To compare red (635 nm) and infrared (780 nm) light for recording pulpal blood flow from human anterior teeth with a laser Doppler flow meter.

DESIGN

Recordings were made from 7 healthy teeth in 5 subjects (aged 22-55 years) using a laser Doppler flow meter (Periflux 4001) equipped with both red and infrared lasers. Average blood flow signals were obtained with both light sources alternately from each tooth under five conditions: intact tooth without opaque rubber dam, intact tooth with dam, after injecting local anaesthetic (3% Mepivacaine) (LA) over the apex of the tooth and cavity preparation to almost expose the pulp, after removal and replacement of the pulp, and with the root canal empty.

RESULTS

With infrared light, because of technical limitations, data were obtained for the first three conditions only. The dam significantly decreased the mean blood flow by 82%. Injecting LA and cavity preparation had no significant effect. With red light, dam produced a decrease of 56%, and the resulting signal was reduced by 33% after LA and cavity preparation. The remaining signal fell by 46% after pulp removal and replacement. This contribution of the pulp is similar to that recorded previously with infrared light. There was no significant further change when the pulp was finally removed.

CONCLUSIONS

The importance of using opaque rubber dam is confirmed. With dam, there is no advantage to using red rather than infrared light, and in each case the pulp contributes less than 50% to the blood flow signal.

摘要

目的

用激光多普勒流量计比较红光(635nm)和红外线(780nm)记录人前牙牙髓血流。

设计

用配备红光和红外线激光的激光多普勒流量计(Periflux 4001)对 5 名受试者(年龄 22-55 岁)的 7 颗健康牙齿进行记录。在 5 种情况下,用两种光源交替对每颗牙齿进行平均血流信号记录:牙齿完整,无不透光橡皮障;牙齿完整,橡皮障覆盖;在牙尖上方注射局部麻醉剂(3%甲哌卡因)(LA);牙体预备至接近暴露牙髓;去除并替换牙髓;根管空。

结果

由于技术限制,用红外线只能获得前三种情况下的数据。橡皮障显著降低平均血流 82%。LA 和牙体预备无显著影响。用红光,橡皮障产生 56%的降低,LA 和牙体预备后信号降低 33%。牙髓去除和替换后,剩余信号降低 46%。这种牙髓的信号与之前用红外线记录的相似。当牙髓最终被去除时,没有进一步的显著变化。

结论

使用不透光橡皮障的重要性得到证实。有橡皮障时,红光与红外线相比没有优势,在每种情况下,牙髓对血流信号的贡献都小于 50%。