Department of Prosthodontics, Faculty of Dentistry, Chiang Mai University, Chiang Mai 50202, Thailand.
Arch Oral Biol. 2013 Nov;58(11):1692-5. doi: 10.1016/j.archoralbio.2013.08.007. Epub 2013 Aug 30.
To determine the effect of enamel and dentine thickness on laser Doppler blood-flow (LDF) signals recorded from dental pulp.
Observations were made on 18 human premolars that had been extracted from young patients during orthodontic treatment. The apical 2/3 of the root was cut off and the remaining pulp removed. Blood flow signals were recorded from the buccal surface of the crown with a laser Doppler flow metre while dilute blood was pumped at 10 ml/min. through a cannula inserted into the pulp cavity. Recordings were made from the enamel surface and at 0.5 mm steps through the enamel and dentine.
The blood flow signal increased significantly as the cavity depth increased and at 2.0 mm, the median flux signal was more than ten times greater than that obtained on the enamel surface. The backscattered light intensity did not change with cavity depth.
When recording pulpal blood flow from a human tooth with a laser Doppler flow metre, a substantially better signal-to-noise ratio should be obtained by placing the probe on dentine in the floor of a cavity than on the enamel surface.
确定牙釉质和牙本质厚度对牙髓激光多普勒血流(LDF)信号的影响。
在正畸治疗中,从年轻患者身上拔出的 18 个人类前磨牙上进行了观察。切除根尖 2/3 的根,去除剩余的牙髓。用激光多普勒流量计从牙冠的颊面记录血流信号,同时通过插入牙髓腔的套管以 10 ml/min 的速度泵送稀释血液。从牙釉质表面以及通过牙釉质和牙本质以 0.5 mm 的步长进行记录。
随着腔深度的增加,血流信号显著增加,在 2.0mm 时,中值通量信号比在牙釉质表面获得的信号大十倍以上。背散射光强度不随腔深度而变化。
用激光多普勒流量计从人牙记录牙髓血流时,将探头置于腔底的牙本质上而不是牙釉质表面上,应该获得更好的信噪比。