Department of Ophthalmology, Miguel Servet University Hospital, Zaragoza, Spain.
Invest Ophthalmol Vis Sci. 2011 Jun 13;52(7):4124-31. doi: 10.1167/iovs.10-6643.
To evaluate the ability of Fourier-domain (FD) optical coherence tomography (OCT) to detect retinal nerve fiber layer (RNFL) atrophy in multiple sclerosis (MS) patients. To test the intrasession reproducibility of RNFL thickness measurements in MS and healthy subjects using Cirrus (Carl Zeiss Meditec, Dublin, CA) and Spectralis (Heidelberg Engineering, Heidelberg, Germany) OCT.
Two hundred twenty-two eyes of 111 subjects (50 MS patients and 61 healthy subjects) underwent three 360° circular scans centered on the optic disc by the same experienced examiner using the Cirrus and Spectralis OCT instruments. Differences between healthy and MS eyes were compared. The relationship between average thicknesses with each OCT was evaluated. Repeatability was studied by intraclass correlation coefficients and coefficients of variation (COV).
RNFL atrophy was detected in the MS eyes for all OCT parameters (P < 0.05). Cirrus and Spectralis showed an RNFL average thickness of 99.4 and 102.5 μm, respectively, in healthy subjects, and 86.0 and 90.4 μm in the MS eyes. RNFL average thickness in the MS eyes determined by both OCTs correlated (r = 0.812; P < 0.001), but were significantly different (P < 0.001). Reproducibility was good. In the MS eyes, Cirrus measurements showed a mean COV of 5.85%, Spectralis 6.80%, and Spectralis with a progression feature 4.16%. Intraclass correlation coefficients were higher than 0.840. RNFL average thickness correlated with disease duration and an optic neuritis antecedent.
There are significant differences in RNFL thickness measurements between Cirrus and Spectralis despite a high correlation of measurement between the two instruments. Fourier-domain OCT can be considered a valid device for detecting RNFL atrophy in MS patients.
评估傅里叶域(FD)光学相干断层扫描(OCT)检测多发性硬化症(MS)患者视网膜神经纤维层(RNFL)萎缩的能力。使用 Cirrus(Carl Zeiss Meditec,都柏林,CA)和 Spectralis(Heidelberg Engineering,海德堡,德国)OCT 测试 MS 和健康受试者的 RNFL 厚度测量的同次内可重复性。
222 只眼 111 例(50 例 MS 患者和 61 例健康受试者),由同一位经验丰富的检查者使用 Cirrus 和 Spectralis OCT 仪器对视盘中心进行三次 360°圆形扫描。比较健康眼和 MS 眼之间的差异。评估每个 OCT 之间的平均厚度关系。通过组内相关系数和变异系数(COV)研究重复性。
所有 OCT 参数均检测到 MS 眼的 RNFL 萎缩(P < 0.05)。Cirrus 和 Spectralis 在健康受试者中分别显示 RNFL 平均厚度为 99.4 和 102.5μm,在 MS 眼中为 86.0 和 90.4μm。两种 OCT 确定的 MS 眼 RNFL 平均厚度相关(r = 0.812;P < 0.001),但存在显著差异(P < 0.001)。重复性好。在 MS 眼中,Cirrus 测量的平均 COV 为 5.85%,Spectralis 为 6.80%,具有进展特征的 Spectralis 为 4.16%。组内相关系数高于 0.840。RNFL 平均厚度与疾病持续时间和视神经炎前体相关。
尽管两种仪器之间的测量相关性较高,但 Cirrus 和 Spectralis 之间的 RNFL 厚度测量存在显著差异。傅里叶域 OCT 可被认为是检测 MS 患者 RNFL 萎缩的有效设备。