Department of Urology, University Hospital, University of Zürich, Zürich, Switzerland.
J Urol. 2011 Jun;185(6):2241-7. doi: 10.1016/j.juro.2011.02.023. Epub 2011 Apr 17.
Technical modifications of the 120 W lithium-triborate laser have been implemented to increase power output, and prevent laser fiber degradation and loss of power output during laser vaporization of the prostate. However, visible alterations at the fiber tip and the subjective impression of decreasing ablative effectiveness during lithium-triborate laser vaporization indicate that delivering constantly high laser power remains a relevant problem. Thus, we evaluated the extent of laser fiber degradation and loss of power output during 120 W lithium-triborate laser vaporization of the prostate.
We investigated 46 laser fibers during routine 120 W lithium-triborate laser vaporization in 35 patients with prostatic bladder outflow obstruction. Laser beam power was measured at baseline and after the application of each 25 kJ during laser vaporization. Fiber tips were microscopically examined after the procedure.
Mild to moderate degradation at the emission window occurred in all fibers, associated with a loss of power output. A steep decrease to a median power output of 57.3% of baseline was detected after applying the first 25 kJ. Median power output at the end of the defined 275 kJ lifespan of the fibers was 48.8%.
Despite technical refinements of the 120 W lithium-triborate laser fiber degradation and significantly decreased power output are still detectable during the procedure. Laser fibers are not fully appropriate for the high power delivery of the new system. There is still potential for further improvement in the laser performance.
对 120W 硼锂酸盐激光器进行了技术改进,以增加功率输出,并防止在前列腺激光汽化过程中激光光纤退化和功率输出损失。然而,在硼锂酸盐激光汽化过程中,光纤尖端出现可见的变化,以及激光消融效果降低的主观印象表明,持续提供高激光功率仍然是一个相关问题。因此,我们评估了在前列腺 120W 硼锂酸盐激光汽化过程中激光光纤退化和功率输出损失的程度。
我们在 35 例前列腺膀胱流出道梗阻患者的常规 120W 硼锂酸盐激光汽化过程中研究了 46 根激光光纤。在激光汽化过程中每次应用 25kJ 时测量激光束功率。在手术结束后,用显微镜检查光纤尖端。
所有光纤的发射窗口都出现了轻度到中度的退化,伴有功率输出损失。在应用前 25kJ 后,检测到功率输出急剧下降至基线的 57.3%。在光纤规定的 275kJ 寿命结束时,功率输出中位数为 48.8%。
尽管对 120W 硼锂酸盐激光光纤进行了技术改进,但在手术过程中仍可检测到光纤退化和功率输出显著下降。激光光纤不完全适合新系统的高功率传输。激光性能仍有进一步提高的潜力。