Jaeken Bart, Lundström Linda, Artal Pablo
Laboratorio de Óptica, Universidad de Murcia, Campus Espinardo (Ed. CiOyN), Murcia, Spain.
Opt Express. 2011 Apr 11;19(8):7903-13. doi: 10.1364/OE.19.007903.
We designed and built a fast scanning peripheral Hartmann-Shack (HS) wave-front sensor to measure the off-axis wave-front aberrations in the human eye. The new instrument is capable of measuring the optical quality over the central 80° horizontal visual field in 1.8 seconds with an angular resolution of 1°. The subject has an open field of view without any moving elements in the line-of-sight and the head is kept in place by a head-chin rest. The same efficiency, reliability and measurement quality as the current static HS sensor were found but with much higher acquisition speed and comfort for the patients. This instrument has the potential to facilitate and improve future research on the peripheral optical quality of the eye in large groups of subjects.
我们设计并制造了一种快速扫描周边哈特曼-夏克(HS)波前传感器,用于测量人眼的离轴波前像差。新仪器能够在1.8秒内测量中央80°水平视野范围内的光学质量,角分辨率为1°。受试者具有开阔的视野,视线中没有任何移动部件,头部通过头托固定在适当位置。该仪器具有与当前静态HS传感器相同的效率、可靠性和测量质量,但采集速度更快,患者舒适度更高。该仪器有潜力促进和改进未来对大量受试者眼睛周边光学质量的研究。