Suppr超能文献

使用激光扫描仪器的超广角周边屈光测量

Ultra-wide-angle peripheral refraction using a laser-scanning instrument.

作者信息

Sager Santiago, Vicente-Jaen Arturo, Lin Zhenghua, Prieto Pedro M, Yang Zhikuang, Lan Weizhong, Artal Pablo

机构信息

Laboratorio de Óptica, Centro de Investigación en Óptica y Nanofísica (CiOyN), Universidad de Murcia, Campus de Espinardo (Ed. 34), 30010 Murcia, Spain.

Voptica S.L., Edificio Pleiades, Universidad de Murcia, Campus de Espinardo (Ed. 37), 30100 Murcia, Spain.

出版信息

Biomed Opt Express. 2024 Oct 24;15(11):6486-6498. doi: 10.1364/BOE.540051. eCollection 2024 Nov 1.

Abstract

We compared the peripheral refractive measurements of a recently proposed laser-scanning instrument with an established peripheral refractor. Two-dimensional refractive maps were obtained using both instruments for 18 young subjects with differing values of central refraction. The comparison shows a strong correlation between devices in the overlapping measurement area, with the new device extending the range of the explored retinal area to a 100-degree-diameter circular patch, compared to the 60°x35° rectangular area of the older peripheral refractor. Larger refractive maps exhibit trends that cannot be easily predicted from narrower scans. These results demonstrate that the new instrument can be a useful tool for assessing wide-angle peripheral optical data in the human eye.

摘要

我们将一种最近提出的激光扫描仪器的周边屈光测量结果与一台已有的周边屈光仪进行了比较。使用这两种仪器为18名具有不同中央屈光值的年轻受试者获取了二维屈光图。比较结果显示,在重叠测量区域内,两种设备之间存在很强的相关性,与旧的周边屈光仪60°×35°的矩形区域相比,新设备将所探索的视网膜区域范围扩展到了直径为100度的圆形区域。更大的屈光图呈现出一些趋势,这些趋势难以从较窄的扫描中轻易预测出来。这些结果表明,新仪器可以成为评估人眼广角周边光学数据的有用工具。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/0434/11563345/2c6cc84c361d/boe-15-11-6486-g001.jpg

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验