Park Young-Woo, Jeong Man-Bok, Kim Tae-Hyun, Ahn Jae-Sang, Ahn Jeong-Taek, Park Shin-Ae, Kim Se-Eun, Seo Kangmoon
Department of Veterinary Surgery and Ophthalmology, College of Veterinary Medicine and BK 21 Program for Veterinary Science, Seoul National University, 599 Gwanak-ro, Gwanak-gu, Seoul 151-742, Korea.
Vet Ophthalmol. 2011 May;14(3):169-73. doi: 10.1111/j.1463-5224.2010.00859.x.
To evaluate the effect of central corneal thickness (CCT) on the measurement of intraocular pressure (IOP) with the rebound (TonoVet(®)) and applanation (TonoPen XL(®)) tonometers in beagle dogs.
Both eyes of 60 clinically normal dogs were used.
The IOP was measured by the TonoVet(®), followed by the TonoPen XL(®) in half of the dogs, while the other half was measured in the reverse order. All CCT measurements were performed 10 min after the use of the second tonometer.
The mean IOP value measured by the TonoVet(®) (16.9 ± 3.7 mmHg) was significantly higher than the TonoPen XL(®) (11.6 ± 2.7 mmHg; P < 0.001). The IOP values obtained by both tonometers were correlated in the regression analysis (γ(2) = 0.4393, P < 0.001). Bland-Altman analysis showed that the lower and upper limits of agreement between the two devices were -0.1 and +10.8 mmHg, respectively. The mean CCT was 549.7 ± 51.0 μm. There was a correlation between the IOP values obtained by the two tonometers and CCT readings in the regression analysis (TonoVet(®) : P = 0.002, TonoPen XL(®) : P = 0.035). The regression equation demonstrated that for every 100 μm increase in CCT, there was an elevation of 1 and 2 mmHg in IOP measured by the TonoPen XL(®) and TonoVet(®), respectively.
The IOP obtained by the TonoVet(®) and TonoPen XL(®) would be affected by variations in the CCT. Therefore, the CCT should be considered when interpreting IOP values measured by tonometers in dogs.
评估中央角膜厚度(CCT)对使用回弹式眼压计(TonoVet®)和压平式眼压计(TonoPen XL®)测量比格犬眼压(IOP)的影响。
使用60只临床健康犬的双眼。
对一半的犬先用TonoVet®测量眼压,再用TonoPen XL®测量;另一半犬则顺序相反。所有CCT测量均在使用第二种眼压计10分钟后进行。
TonoVet®测量的平均眼压值(16.9±3.7 mmHg)显著高于TonoPen XL®(11.6±2.7 mmHg;P<0.001)。回归分析显示两种眼压计测得的眼压值具有相关性(γ² = 0.4393,P<0.001)。Bland-Altman分析表明,两种设备之间的一致性下限和上限分别为-0.1和+10.8 mmHg。平均CCT为549.7±51.0μm。回归分析表明,两种眼压计测得的眼压值与CCT读数之间存在相关性(TonoVet®:P = 0.002,TonoPen XL®:P = 0.035)。回归方程表明,CCT每增加100μm,TonoPen XL®和TonoVet®测量的眼压分别升高1 mmHg和2 mmHg。