Allen J K, Hensley W J, Nicholls A V, Whitfield J B
Clin Chem. 1979 Feb;25(2):325-7.
Enzymic measurement of high-density lipoprotein cholesterol with a centrifugal analyzer is described. We used polyethylene glycol (Mr 6000), final concentration 100 g/L, to precipitate low-density and very-low-density lipoproteins, thereby eliminating the difficulties of the commonly used heparin/Mn2+ precipitation method and facilitating the use of ethylenediaminetetraacetate-stabilized plasma. As measured by rocket immunoelectrophoresis, this final concentration of polyethylene glycol completely precipitates beta-lipoproteins, leaving the alpha-lipoproteins in solution. Between-run reproducibility (CV) was 3.6%, within-run reproducibility (CV) 0.8%. Reagent costs currently are $US 0.13 per test and large numbers of samples can be handled conveniently. Normal ranges were compiled for 539 men and 444 women. The high-density lipoprotein cholesterol for men was 1.20 +/- 0.31 (SD) mmol/L and for women 1.52 +/- 0.38 (SD) mmol/L.
本文描述了用离心分析仪酶法测定高密度脂蛋白胆固醇的方法。我们使用终浓度为100 g/L的聚乙二醇(分子量6000)沉淀低密度和极低密度脂蛋白,从而消除了常用的肝素/Mn2+沉淀法的困难,并便于使用乙二胺四乙酸稳定的血浆。通过火箭免疫电泳测定,该终浓度的聚乙二醇能完全沉淀β-脂蛋白,而α-脂蛋白则留在溶液中。批间重复性(CV)为3.6%,批内重复性(CV)为0.8%。目前试剂成本为每次检测0.13美元,且能方便地处理大量样本。编制了539名男性和444名女性的正常范围。男性的高密度脂蛋白胆固醇为1.20±0.31(标准差)mmol/L,女性为1.52±0.38(标准差)mmol/L。