Warnick G R, Cheung M C, Albers J J
Clin Chem. 1979 Apr;25(4):596-604.
We compared six precipitation methods for high-density-lipoprotein cholesterol quantitation with an ultracentrifugation method, the accuracy of which was improved by correcting for 4% manipulative loss in the d greater than 1.063 fractions. For purposes of comparison, the apoprotein B-associated cholesterol (average 56 mg/L) measured by immunoassay in the d greater than 1.063 fractions was subtracted. In 65 plasma samples from men, women, and children, a heparin-Mn2+ procedure with Mn2+ at 46 mmol/L produced results slightly higher (+16 mg/L), while results with Mn2+ at 92 mmol/L averaged slightly lower (-8 mg/L) than the comparison ultracentrifuge method. Results that were about 5% low were obtained by the dextran sulfate 500-Mg2+ (-25 mg/L) and phosphotungstate-Mg2+ (-31 mg/L) methods. A heparin-Ca2+ method produced results 10% high (+58 mg/L). Results by a polyethylene glycol-6000 precipitation method were 12% low (-64 mg/L). Precision was better with the two heparin-Mn2+ and the dextran sulfate 500-Mg2+ procedures, with CVs of 4%, intermediate with phosphotungstate-Mg2+ and polyethylene glycol-6000 (CV 6-7%), and poorest with heparin-Ca2+ (CV 10%). Precipitation by phosphotungate-Mg2+ appeared more sensitive to reagent concentration and temperature variations than either the heparin-Mn2+ or dextran sulfate 500-Mg2+ methods. We conclude that these precipitation methods are not equivalent, but give rise to significant systematic differences in high-density-lipoprotein cholesterol quantitation.
我们将六种用于高密度脂蛋白胆固醇定量的沉淀法与一种超速离心法进行了比较,通过校正d大于1.063组分中4%的操作损失提高了超速离心法的准确性。为了进行比较,减去了通过免疫测定法在d大于1.063组分中测得的载脂蛋白B相关胆固醇(平均56mg/L)。在来自男性、女性和儿童的65份血浆样本中,46mmol/L Mn2+的肝素-Mn2+法产生的结果略高(+16mg/L),而92mmol/L Mn2+的结果平均略低于比较超速离心法(-8mg/L)。硫酸葡聚糖500-Mg2+法(-25mg/L)和磷钨酸盐-Mg2+法(-31mg/L)得到的结果低约5%。肝素-Ca2+法产生的结果高10%(+58mg/L)。聚乙二醇6000沉淀法的结果低12%(-64mg/L)。两种肝素-Mn2+法和硫酸葡聚糖500-Mg2+法的精密度较好,变异系数为4%,磷钨酸盐-Mg2+法和聚乙二醇6000法的精密度中等(变异系数6-7%),肝素-Ca2+法最差(变异系数10%)。与肝素-Mn2+法或硫酸葡聚糖500-Mg2+法相比,磷钨酸盐-Mg2+沉淀法似乎对试剂浓度和温度变化更敏感。我们得出结论,这些沉淀法并不等效,而是在高密度脂蛋白胆固醇定量中产生显著的系统差异。