OrthoCarolina, PA, Charlotte, North Carolina 28207, USA.
J Arthroplasty. 2011 Sep;26(6 Suppl):14-8. doi: 10.1016/j.arth.2011.03.027. Epub 2011 May 8.
The ASR (articular surface replacement) XL (DePuy, Warsaw, Ind) metal-on-metal hip arthroplasty offers the advantage of stability and increased motion. However, an alarming number of early failures prompted the evaluation of patients treated with this system. A prospective study of patients who underwent arthroplasty with the ASR XL system was performed. Patients with 2-year follow-up or any revision were included. Failure rates, Western Ontario and McMaster Universities Osteoarthritis Index (WOMAC) scores, and radiographs were evaluated. Ninety-five patients (105 hips) were included. There were 16 revisions. Thirteen (12%) were aseptic acetabular failures. Eight were revised for aseptic loosening; 4, for metallosis; 1, for malposition; 2, for infection; and 1, for periprosthetic fracture. Mean time to revision was 1.6 years (0.18-3.4 years). The ASR XL with a revision rate of 12% is the second reported 1 piece metal-on-metal system with a significant failure rate at early follow-up. This particular class of implants has inherent design flaws that lead to early failure.
ASR(关节表面置换)XL(DePuy,印第安纳州华沙)金属对金属髋关节置换术具有稳定性和增加活动度的优势。然而,大量早期失败促使人们对接受该系统治疗的患者进行评估。对接受 ASR XL 系统关节置换术的患者进行了前瞻性研究。纳入了有 2 年随访或任何翻修的患者。评估了失败率、西安大略和麦克马斯特大学骨关节炎指数(WOMAC)评分和 X 线片。共纳入 95 例患者(105 髋),有 16 例翻修。13 例(12%)为无菌性髋臼失败。8 例因无菌性松动而翻修;4 例因金属病而翻修;1 例因位置不当而翻修;2 例因感染而翻修;1 例因假体周围骨折而翻修。平均翻修时间为 1.6 年(0.18-3.4 年)。ASR XL 的翻修率为 12%,是第二个报道的在早期随访中有显著失败率的 1 件式金属对金属系统。这类植入物具有内在的设计缺陷,导致早期失败。