Bos I, Lindner B, Seydel U, Johannisson R, Dörre E, Henssge J, Löhrs U
Institut für Pathologie, Medizinische Universität zu Lübeck.
Z Orthop Ihre Grenzgeb. 1990 Jan-Feb;128(1):73-82. doi: 10.1055/s-2008-1039865.
Aseptic loosening of joint prostheses is a quantitatively increasing problem. For evaluation of the pathogenesis of prosthesis loosening joint capsules and tissue membranes taken from the femoral and acetabular bone-cement-interface of 23 hips revised for aseptic loosening were examined. In all cases we found an excessive inflammatory histiocytic reaction with intracytoplasmatic incorporation of a small granular foreign material. This could be identified to be zirconium oxide by LAMMA-analysis. Zirconium oxide and polymethylmethacrylate (PMMA) wear particles could be shown in the cytoplasm of histiocytes by transmission electron microscopy. By quantitative grading of the histiocytic reaction and the intracellular granular wear particles we found that both were evenly distributed around the loosened prosthesis. Chronic inflammation due to bone cement abrasion is regarded as a leading cause of aseptic prosthesis loosening.
关节假体的无菌性松动是一个数量上不断增加的问题。为了评估假体松动的发病机制,对23例因无菌性松动而翻修的髋关节的关节囊和取自股骨及髋臼骨水泥界面的组织膜进行了检查。在所有病例中,我们发现了过度的炎症性组织细胞反应,胞质内含有小颗粒状异物。通过激光微探针质谱分析(LAMMA分析)可确定其为氧化锆。通过透射电子显微镜可在组织细胞的细胞质中显示出氧化锆和聚甲基丙烯酸甲酯(PMMA)磨损颗粒。通过对组织细胞反应和细胞内颗粒状磨损颗粒进行定量分级,我们发现两者在松动的假体周围分布均匀。骨水泥磨损导致的慢性炎症被认为是无菌性假体松动的主要原因。