Folks Russell D, Manatunga Daya, Garcia Ernest V, Taylor Andrew T
Department of Radiology, School of Medicine, Emory University, Atlanta, Georgia, USA.
J Nucl Med Technol. 2011 Jun;39(2):131-9. doi: 10.2967/jnmt.110.081893. Epub 2011 May 12.
Kidney motion during dynamic renal scintigraphy can cause errors in calculated renal function parameters. Our goal was to develop and validate algorithms to detect and correct patient motion.
We retrospectively collected dynamic images from 86 clinical renal studies (42 women, 44 men), acquired using (99m)Tc-mercaptoacetyltriglycine (80 image frames [128 × 128 pixels; 3.2 mm/pixel]: twenty-four 2-s frames, sixteen 15-s frames, and forty 30-s frames). We simulated 10 types of vertical motion in each patient study, resulting in 860 image sets. Motion consisted of up or down shifts of magnitude 0.25 pixel to 4 pixels per frame and was either a gradual shift additive over multiple frames or an abrupt shift of one or more consecutive frames, with a later return to the start position. Additional horizontal motion was added to test its effect on detection of vertical motion. Original and shifted files were processed using a motion detection algorithm. Corrective shifts were applied, and the corrected and original (unshifted) images were compared pixel by pixel. Motion detected in the shifted data was also tabulated before and after correction of motion detected in the original data. A detected shift was considered correct if it was within 0.25 pixel of the simulated magnitude. Software was developed to facilitate visual review of all images and to summarize kidney motion and motion correction using linograms.
Overall detection of simulated shifts was 99% (3,068/3,096 frames) when the existing motion in the original images was first corrected. When the original motion was not corrected, overall shift detection was 76% (2,345/3,096 frames). For image frames in which no shift was added (and original motion was not corrected), 87% (27,142/31,132 frames) were correctly detected as having no shift. When corrected images were compared with original images, calculated count recovery was 100% for all shifts that were whole-pixel magnitudes. For fractional-pixel shifts, percentage count recovery varied from 52% to 73%. Visual review suggested that some original frames exhibited true patient motion.
The algorithm accurately detected motion as small as 0.25 pixel. Whole-pixel motion can be detected and corrected with high accuracy. Fractional-pixel motion can be detected and corrected, but with less accuracy. Importantly, by accurately identifying unshifted frames, the algorithm helps to prevent the introduction of errors during motion correction.
动态肾闪烁显像期间肾脏运动可导致计算出的肾功能参数出现误差。我们的目标是开发并验证用于检测和校正患者运动的算法。
我们回顾性收集了86例临床肾脏研究的动态图像(42例女性,44例男性),这些图像使用(99m)锝-巯基乙酰三甘氨酸采集(80个图像帧[128×128像素;3.2毫米/像素]:24个2秒帧、16个15秒帧和40个30秒帧)。我们在每项患者研究中模拟了10种垂直运动类型,从而得到860个图像集。运动包括每帧大小为0.25像素至4像素的向上或向下偏移,并且要么是在多个帧上逐渐累加的偏移,要么是一个或多个连续帧的突然偏移,随后返回起始位置。还添加了额外的水平运动以测试其对垂直运动检测的影响。使用运动检测算法对原始文件和偏移后的文件进行处理。应用校正偏移,然后逐像素比较校正后的图像和原始(未偏移)图像。在原始数据中检测到的运动校正前后,还将偏移数据中检测到的运动制成表格。如果检测到的偏移在模拟大小的0.25像素范围内,则认为是正确的。开发了软件以方便对所有图像进行视觉检查,并使用线性图总结肾脏运动和运动校正情况。
当首先校正原始图像中的现有运动时,模拟偏移的总体检测率为99%(3068/3096帧)。当未校正原始运动时,总体偏移检测率为76%(2345/3096帧)。对于未添加偏移(且未校正原始运动)的图像帧,87%(27142/31132帧)被正确检测为无偏移。当将校正后的图像与原始图像进行比较时,对于所有整像素大小的偏移,计算出的计数恢复率为100%。对于分数像素偏移,计数恢复百分比从52%到73%不等。视觉检查表明一些原始帧显示出真正的患者运动。
该算法能准确检测小至0.25像素的运动。整像素运动能够被高精度地检测和校正。分数像素运动能够被检测和校正,但精度较低。重要的是,通过准确识别未偏移的帧,该算法有助于防止在运动校正期间引入误差。