Baillargeon Pierre, Scampavia Louis, Einsteder Ross, Hodder Peter
Lead Identification Division, Translational Research Institute, Scripps Research Institute Molecular Screening Center, Jupiter, FL 33458, USA.
J Lab Autom. 2011 Jun;16(3):197-203. doi: 10.1016/j.jala.2011.02.004. Epub 2011 May 16.
This report presents the high-resolution image acquisition and processing instrument for compound management applications (HIAPI-CM). The HIAPI-CM combines imaging spectroscopy and machine-vision analysis to perform rapid assessment of high-throughput screening (HTS) compound library quality. It has been customized to detect and classify typical artifacts found in HTS compound library microtiter plates (MTPs). These artifacts include (1) insufficient volume of liquid compound sample, (2) compound precipitation, and (3) colored compounds that interfere with HTS assay detection format readout. The HIAPI-CM is also configured to automatically query and compare its analysis results to data stored in a LIMS or corporate database, aiding in the detection of compound registration errors. To demonstrate its capabilities, several compound plates (n=5760 wells total) containing different artifacts were measured via automated HIAPI-CM analysis, and results compared with those obtained by manual (visual) inspection. In all cases, the instrument demonstrated high fidelity (99.8% empty wells; 100.1% filled wells; 94.4% for partially filled wells; 94.0% for wells containing colored compounds), and in the case of precipitate detection, the HIAPI-CM results significantly exceeded the fidelity of visual observations (220.0%). As described, the HIAPI-CM allows for noninvasive, nondestructive MTP assessment with a diagnostic throughput of about 1min per plate, reducing analytical expenses and improving the quality and stewardship of HTS compound libraries.
本报告介绍了用于化合物管理应用的高分辨率图像采集与处理仪器(HIAPI-CM)。HIAPI-CM结合成像光谱和机器视觉分析,对高通量筛选(HTS)化合物库质量进行快速评估。它经过定制,可检测和分类HTS化合物库微孔板(MTP)中发现的典型伪像。这些伪像包括:(1)液体化合物样品体积不足;(2)化合物沉淀;(3)干扰HTS分析检测形式读数的有色化合物。HIAPI-CM还配置为自动查询其分析结果,并与存储在实验室信息管理系统(LIMS)或公司数据库中的数据进行比较,有助于检测化合物登记错误。为展示其功能,通过HIAPI-CM自动分析对几块含有不同伪像的化合物板(共5760孔)进行了测量,并将结果与手动(目视)检查获得的结果进行比较。在所有情况下,该仪器都显示出高保真度(空孔为99.8%;填充孔为100.1%;部分填充孔为94.4%;含有有色化合物的孔为94.0%),在沉淀检测方面,HIAPI-CM的结果显著超过目视观察的保真度(220.0%)。如所述,HIAPI-CM允许对MTP进行非侵入性、无损评估,诊断通量约为每板1分钟,降低了分析成本,提高了HTS化合物库的质量和管理水平。