We report a case of a 46-year-old male patient who sustained a fracture of the Burch-Schneider anti-protrusio cage (APC) that was implanted 13 years before. The fracture occurred in the transitional area to the proximal flange and was caused by direct and recurrent contact between the prosthetic metal head and the acetabular device. Late failure was related to increased wear of the polyethylene cup further aggravated by vertical orientation of the cage. Revision surgery included acetabular reconstruction using a new APC and cross-linked polyethylene cup, and replacement of the 36 mm diameter ball head. At 18 months follow-up the patient showed a complete functional recovery. When a reconstruction device is used, particularly in a young patient, special attention should be paid to the correct positioning of the cage in order to prevent accelerated polyethylene wear. Furthermore, cross-linked polyethylene cups that have become available in recent years must be used.
我们报告一例46岁男性患者,其13年前植入的Burch-Schneider防髋臼前突笼(APC)发生骨折。骨折发生在靠近近端凸缘的过渡区域,是由假体金属头与髋臼装置之间的直接反复接触所致。晚期失败与聚乙烯杯磨损增加有关,而笼的垂直方向进一步加剧了这种磨损。翻修手术包括使用新的APC和交联聚乙烯杯进行髋臼重建,以及更换直径36 mm的球头。在18个月的随访中,患者功能完全恢复。当使用重建装置时,尤其是在年轻患者中,应特别注意笼的正确定位,以防止聚乙烯加速磨损。此外,必须使用近年来已有的交联聚乙烯杯。