维生素 E 注入的超高分子量聚乙烯髋臼杯的磨损和迁移不受头大小的影响。

Wear and migration are not influenced by head size in a vitamin E-infused highly cross-linked polyethylene acetabular cup.

机构信息

Mathys Ltd Bettlach, Robert-Mathys Strasse 5, 2544 Bettlach, Switzerland.

Diakonessenhuis Utrecht, Bosboomstraat 1, 3582 KE Utrecht, Netherlands; Clinical Orthopedic Research Centre midden-Nederland, Diakonessenhuis Zeist, Jagersingel 1, 3707 HL Zeist, Netherlands.

出版信息

Orthop Traumatol Surg Res. 2021 Feb;107(1):102644. doi: 10.1016/j.otsr.2020.03.030. Epub 2020 Dec 28.

Abstract

BACKGROUND

Aseptic loosening and periprosthetic osteolysis are frequent complications in total hip arthroplasty requiring revision surgery. Highly cross-linked polyethylene (HXLPE) implants have improved wear resistance, permitting larger femoral heads. However, such implants may experience surface cracking, mechanical failure, and oxidative damage. Vitamin E-infused HXLPE (VEPE) implants were therefore developed to reduce oxidation without compromising mechanical strength. We addressed the following questions: (1) Does femoral head size affect the midterm annual polyethylene wear rates of VEPE acetabular cups? (2) Does femoral head size affect the midterm migration rates of VEPE acetabular cups? (3) Are clinical outcomes affected by femoral head size?

HYPOTHESIS

Annual wear rate, migration rate, and clinical outcomes of VEPE acetabular cups are independent of femoral head size.

PATIENTS AND METHODS

This was a prospective, multicentre, observational study of patients that underwent total hip arthroplasty. Hips were grouped according to the size of femoral head implanted (28 mm, 32 mm, and 36 mm). We determined annual wear rate and migration rate of VEPE acetabular cups using the Einzel-Bild-Röntgen-Analyse software. Clinically, we assessed the Harris Hip Score and visual analog score for pain and satisfaction.

RESULTS

We followed 253 patients (267 hips) for a mean of 55.0±20.6 months in the 28 mm, 46.2±21.4 months in the 32 mm, and 43.8±22.6 months in the 36 mm group. The annual wear rate was 0.025 mm per year from 1 year to the last follow-up and remained similar between the groups (p>0.05). Also, mean two-dimensional migration rates did not exceed 0.05 mm from 2 years to the last follow-up and remained similar between the groups (p=0.355). Finally, clinical outcomes also did not differ between the groups (p>0.05). Two patients required revision surgery.

DISCUSSION

Femoral head size did not influence midterm annual wear rate, migration rate, and clinical outcomes of VEPE acetabular cups. Furthermore, wear and cup migration rates were below the reported values leading to osteolysis and aseptic loosening. Nevertheless, studies with extended follow-up periods will be necessary to confirm these results in the long term.

LEVEL OF EVIDENCE

IV.

摘要

背景

在全髋关节置换术中,无菌性松动和假体周围骨溶解是常见的并发症,需要进行翻修手术。高度交联聚乙烯(HXLPE)植入物提高了耐磨性,允许使用更大的股骨头。然而,此类植入物可能会出现表面裂纹、机械故障和氧化损伤。因此,开发了维生素 E 注入的 HXLPE(VEPE)植入物以减少氧化而不影响机械强度。我们提出了以下问题:(1)股骨头大小会影响 VEPE 髋臼杯的中期年度聚乙烯磨损率吗?(2)股骨头大小会影响 VEPE 髋臼杯的中期迁移率吗?(3)临床结果是否受股骨头大小的影响?

假设

VEPE 髋臼杯的年磨损率、迁移率和临床结果与股骨头大小无关。

患者和方法

这是一项前瞻性、多中心、观察性研究,纳入了接受全髋关节置换术的患者。根据植入的股骨头大小(28mm、32mm 和 36mm)将髋关节分组。我们使用 Einzel-Bild-Röntgen-Analyse 软件确定 VEPE 髋臼杯的年磨损率和迁移率。临床方面,我们评估了 Harris 髋关节评分以及疼痛和满意度的视觉模拟评分。

结果

我们随访了 253 名患者(267 髋),平均随访时间为 28mm 组 55.0±20.6 个月,32mm 组 46.2±21.4 个月,36mm 组 43.8±22.6 个月。从 1 年到最后一次随访,年磨损率为 0.025mm/年,各组之间相似(p>0.05)。同样,从 2 年到最后一次随访,二维平均迁移率也不超过 0.05mm,各组之间相似(p=0.355)。最后,各组之间的临床结果也没有差异(p>0.05)。有 2 名患者需要翻修手术。

讨论

股骨头大小不影响 VEPE 髋臼杯的中期年度磨损率、迁移率和临床结果。此外,磨损和杯迁移率低于导致骨溶解和无菌性松动的报告值。然而,需要进行延长随访期的研究来长期证实这些结果。

证据水平

IV。