Performance Trade-Off Analysis Comparing Different Front-End Configurations for a Digital X-ray Imager.

作者信息

Kuhls-Gilcrist Andrew, Jain Amit, Bednarek Daniel R, Rudin Stephen

机构信息

A. Kuhls-Gilcrist was with the Toshiba Stroke Research Center, University at Buffalo, Buffalo, NY 14214 USA. He is now with Toshiba America Medical Systems, Inc., Tustin, CA 92780 USA (telephone: 714-669-7694,

出版信息

IEEE Nucl Sci Symp Conf Rec (1997). 2010 Oct 30;2010:2491-2494. doi: 10.1109/NSSMIC.2010.5874235.

Abstract

Performance of indirect digital x-ray imagers is typically limited by the front-end components. Present x-ray-to-light converting phosphors significantly reduce detector resolution due to stochastic blurring and k-fluorescent x-ray reabsorption. Thinner phosphors improve resolution at the cost of lowering quantum detection efficiency (QDE) and increasing Swank noise. Magnifying fiber optic tapers (FOTs) are commonly used to increase the field-of-view of small sensor imagers, such as CMOS, CCD, or electron-multiplying CCD (EMCCD) based detectors, which results in a reduction in detector sensitivity and further reduces the MTF. We investigate performance trade-offs for different front-end configurations coupled to an EMCCD sensor with 8 μm pixels. Six different columnar structured CsI(Tl) scintillators with thicknesses of 100, 200, 350, 500, and 1000 μm type high-light (HL) and a 350 μm type high-resolution (HR) (Hamamatsu) and four different FOTs with magnification ratios (M) of 1, 2.5, 3.3, and 4 were studied using the RQA5 x-ray spectrum. The relative signal of the different scintillators largely followed the relative QDE, indicating their light output per absorbed x-ray was similar, with the type HR CsI emitting 57% of the type HL. The efficiency of the FOTs was inversely proportional to M(2) with the M = 1 FOT transmitting 87% of the incident light. At 5 (10) cycles/mm, the CsI MTF was 0.38 (0.22), 0.33 (0.17), 0.37 (0.19), 0.23 (0.09), 0.19 (0.08), and 0.09 (0.03) for the 100, 200, 350HR, 350, 500, and 1000 μm CsI, respectively and the FOT MTF was 0.89 (0.84), 0.80 (0.72), 0.70 (0.60), and 0.69 (0.37) for M = 1, 2.5, 3.3, and 4, respectively. The 1000, 500, and 350HR μm CsI had the highest DQE for low, medium, and high spatial frequency ranges of 0 to 1.6, 1.6 to 4.5, and 4.5 to 10 cycles/mm, respectively. Larger FOT M resulted in a reduction in DQE. Quantifying performance of different front-end configurations will enable optimal selection of components for task-specific designs.

摘要

相似文献

1
Performance Trade-Off Analysis Comparing Different Front-End Configurations for a Digital X-ray Imager.
IEEE Nucl Sci Symp Conf Rec (1997). 2010 Oct 30;2010:2491-2494. doi: 10.1109/NSSMIC.2010.5874235.
2
X-ray imaging performance of structured cesium iodide scintillators.
Med Phys. 2004 Sep;31(9):2594-605. doi: 10.1118/1.1782676.
10
The Solid-State X-Ray Image Intensifier (SSXII): An EMCCD-Based X-Ray Detector.
Proc SPIE Int Soc Opt Eng. 2008;6913:69130K. doi: 10.1117/12.772724.

引用本文的文献

本文引用的文献

2
The Solid State X-ray Image Intensifier (SSXII) in Single Photon Counting (SPC) mode.
Proc SPIE Int Soc Opt Eng. 2010 Mar 22;7622(76221P). doi: 10.1117/12.843785.
3
A method for the determination of the two-dimensional MTF of digital radiography systems using only the noise response.
Proc SPIE Int Soc Opt Eng. 2010 Mar 23;7622(76224W):76224W-76244W9. doi: 10.1117/12.843918.
4
Accurate MTF measurement in digital radiography using noise response.
Med Phys. 2010 Feb;37(2):724-35. doi: 10.1118/1.3284376.
6
The Solid-State X-Ray Image Intensifier (SSXII): An EMCCD-Based X-Ray Detector.
Proc SPIE Int Soc Opt Eng. 2008;6913:69130K. doi: 10.1117/12.772724.
7
X-ray imaging performance of structured cesium iodide scintillators.
Med Phys. 2004 Sep;31(9):2594-605. doi: 10.1118/1.1782676.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验