AJR Am J Roentgenol. 2011 Aug;197(2):468-73. doi: 10.2214/AJR.10.6058.
There are known interoperator, intraoperator, and intervendor software differences that can influence the reproducibility of quantitative CT perfusion values. The purpose of this study was to determine the relative impact of operator and software differences in CT perfusion variability.
CT perfusion imaging data were selected for 11 patients evaluated for suspected ischemic stroke. Three radiologists each independently postprocessed the source data twice, using four different vendor software applications. Results for cerebral blood volume (CBV), cerebral blood flow (CBF), and mean transit time (MTT) were recorded for the lentiform nuclei in both hemispheres. Repeated variables multivariate analysis of variance was used to assess differences in the means of CBV, CBF, and MTT. Bland-Altman analysis was used to assess agreement between pairs of vendors, readers, and read times.
Choice of vendor software, but not interoperator or intraoperator disagreement, was associated with significant variability (p < 0.001) in CBV, CBF, and MTT. The mean difference in CT perfusion values was greater for pairs of vendors than for pairs of operators.
Different vendor software applications do not generate quantitative perfusion results equivalently. Intervendor difference is, by far, the largest cause of variability in perfusion results relative to interoperator and intraoperator difference. Caution should be exercised when interpreting quantitative CT perfusion results because these values may vary considerably depending on the postprocessing software.
存在已知的操作者间、操作者内和供应商间软件差异,这些差异可能会影响定量 CT 灌注值的可重复性。本研究旨在确定操作者和软件差异对 CT 灌注变异性的相对影响。
选择了 11 例疑似缺血性卒中患者的 CT 灌注成像数据。3 名放射科医生分别独立地使用 4 种不同供应商的软件应用程序,对原始数据进行了两次后处理。记录双侧豆状核的脑血容量(CBV)、脑血流量(CBF)和平均通过时间(MTT)的结果。采用重复变量方差分析评估 CBV、CBF 和 MTT 的均值差异。Bland-Altman 分析用于评估供应商、读者和阅读次数之间的一致性。
供应商软件的选择,但不是操作者间或操作者内的分歧,与 CBV、CBF 和 MTT 的显著变异性(p<0.001)相关。供应商对之间的 CT 灌注值的平均差异大于操作者对之间的差异。
不同的供应商软件应用程序不能等效地生成定量灌注结果。与操作者间和操作者内差异相比,供应商间差异是造成灌注结果变异性的最大原因。在解释定量 CT 灌注结果时应谨慎,因为这些值可能会因后处理软件的不同而有很大差异。