Department of Endodontics and Restorative Dentistry, University of Siena, Policlinico Le Scotte, Viale Bracci, Siena, Italy.
J Endod. 2011 Jun;37(6):847-50. doi: 10.1016/j.joen.2011.02.030. Epub 2011 Apr 6.
The shaping capacity of nickel-titanium (NiTi) rotary instruments is often assessed by photographic or micro-computed tomography (micro-CT) measurements, and these instruments are often used more than once clinically. This study was conducted to compare photographic and micro-CT measurements and to assess if the repeated use of NiTi instruments affected the shape of canal preparation.
Ten new sets of ProTaper Universal instruments (Dentsply-Maillefer, Ballaigues, Switzerland) were used in 60 resin blocks simulating curved root canals. Groups 1 to 6 (n=10) represented the first to sixth use of the instrument, respectively. Digitized images of the prepared blocks were taken in both mesiodistal (MD) and buccolingual (BL) directions and area measurements (mm(2)) were calculated using AutoCAD (Autodesk Inc, San Rafael, CA). The volumes of the same prepared canals were measured using micro-CT (mm(3)). Statistical analysis was performed to detect differences between photographic and volumetric measurements and differences between uses.
Two-way repeated-measures analysis of variance revealed significant differences between groups (P < .001). Regarding measurement type, there were no significant differences between BL and MD measurements, but there were significant differences between micro-CT and BL measurements (P < .001) and micro-CT and MD measurements (P=.001). Significant differences were also noted between uses.
Within the limitations of the present study, micro-CT scanning is more discriminative of the changes in canal space associated with repeated instrument use than photographic measurements. Canal preparations are significantly smaller after the third use of the same instrument.
镍钛(NiTi)旋转器械的成形能力通常通过摄影或微计算机断层扫描(micro-CT)测量来评估,并且这些器械在临床上通常会被多次使用。本研究旨在比较摄影和 micro-CT 测量,并评估 NiTi 器械的重复使用是否会影响根管预备的形状。
10 套新的 ProTaper Universal 器械(Dentsply-Maillefer,Ballaigues,瑞士)用于模拟弯曲根管的 60 个树脂块。第 1 至 6 组(n=10)分别代表器械的第 1 至第 6 次使用。使用 AutoCAD(Autodesk Inc,San Rafael,CA)对预备块的近远中(MD)和颊舌向(BL)数字化图像进行拍摄,并计算面积测量值(mm(2))。使用 micro-CT(mm(3))测量相同预备根管的体积。采用双向重复测量方差分析检测组间、测量类型间和使用次数间的差异。
双向重复测量方差分析显示组间差异有统计学意义(P <.001)。关于测量类型,BL 和 MD 测量值之间无显著差异,但 micro-CT 和 BL 测量值(P <.001)以及 micro-CT 和 MD 测量值(P=.001)之间存在显著差异。使用次数间也存在显著差异。
在本研究的限制范围内,micro-CT 扫描比摄影测量更能区分与器械重复使用相关的根管空间变化。同一器械使用第 3 次后,根管预备明显变小。