Department of Emergency Medicine, Henry Ford Health, System, Detroit, MI 48202, USA.
Am J Emerg Med. 2011 Sep;29(7):782-9. doi: 10.1016/j.ajem.2011.05.014. Epub 2011 Jul 29.
Continuous invasive blood pressure (BP) and heart rate (HR) monitoring in the emergency department (ED) is valuable in managing critically ill patients. Novel noninvasive finger cuff technology allows this same uninterrupted monitoring for almost any individual. This exploratory study compares ED noninvasive continuous to intermittent measurements of these variables.
A variety of acutely ill ED patients (n = 40) with broad ranges of BP and HR underwent simultaneous monitoring using interval standard automated ED devices and continuous finger cuff technology (Nexfin; Bmeye, Amsterdam, The Netherlands) over a period of 2 hours. At baseline and at 15-minute intervals, simultaneous measurements for BP and HR were recorded and compared.
There were 25 men and 15 women enrolled with a mean age of 62.2 years (SD, 12.6). Eighteen had acute dyspnea, 11 with probable stroke syndrome, 3 with suspected sepsis, and 8 with a systolic BP greater than 180 or less than 100 mm Hg. Blood pressure and HR values (n = 344) simultaneously measured by automated ED equipment and the Nexfin finger cuff device were compared. The Pearson correlation coefficients were high, and the Bland-Altman plots showed good agreement between the 2 determinations.
Continuous BP and HR monitoring measured by the Nexfin finger cuff device in this trial showed reasonable agreement when compared with the intermittent values obtained by automated ED equipment. However, theoretically, noninvasive and continuous monitoring of the BP and HR might better reflect underlying hemodynamics than these same measurements obtained intermittently and, thus, could be important in patient management. More study is needed to determine the optimal method of monitoring these parameters.
在急诊科(ED)持续进行有创血压(BP)和心率(HR)监测对于危重症患者的管理具有重要价值。新型非侵入性手指袖带技术几乎可以对任何个体进行同样的不间断监测。本探索性研究比较了 ED 非侵入性连续与这些变量的间歇性测量。
在 2 小时的时间内,对各种患有广泛 BP 和 HR 范围的急性 ED 患者(n = 40)同时使用间隔标准自动化 ED 设备和连续手指袖带技术(Nexfin;Bmeye,阿姆斯特丹,荷兰)进行监测。在基线和每 15 分钟间隔时,记录并比较同时进行的 BP 和 HR 测量值。
共纳入 25 名男性和 15 名女性,平均年龄为 62.2 岁(SD,12.6)。18 人有急性呼吸困难,11 人有疑似中风综合征,3 人有疑似脓毒症,8 人有收缩压大于 180 或小于 100mmHg。同时使用自动化 ED 设备和 Nexfin 手指袖带设备测量血压和 HR 值(n = 344)。Pearson 相关系数较高,Bland-Altman 图显示两种测定方法之间具有良好的一致性。
在本试验中,与自动化 ED 设备获得的间歇性值相比,Nexfin 手指袖带设备测量的连续 BP 和 HR 监测具有较好的一致性。然而,从理论上讲,BP 和 HR 的非侵入性和连续监测可能比这些相同的间歇性测量更能反映潜在的血流动力学,因此在患者管理中可能很重要。需要进一步研究以确定监测这些参数的最佳方法。