Cytometry. 1990;11(2):223-30. doi: 10.1002/cyto.990110202.
The epi-illumination optics employed in most arc lamp-based flow cytometers may be modified so as to produce a dark-field configuration which facilitates highly sensitive detection of both forward and large angle light scattering in an instrument with a "jet on open surface" flow chamber. Forward scattering is detected at angles upwards from about 2 degrees, while large angle scattering includes angles above 18 degrees. Theoretical considerations suggest that large angle scattering measured around 20 degrees may be as efficient as that measured at 90 degrees for the purpose of distinguishing cells on the basis of intracellular structure. This was supported by the finding that dual parameter light scattering histograms of leukocyte suspensions obtained with the arc lamp-based instrument were closely similar to such histograms recorded with a laser-based instrument with the large angle detector at 90 degrees. Different species of bacteria could be distinguished by means of the dual parameter light scattering device, as could different species of sea algae. The sensitivity of the device is sufficient to measure 0.2 microns polystyrene particles in both forward and large angle scattering.
大多数基于弧光灯的流式细胞仪所采用的落射照明光学系统可以进行改进,以产生暗场配置,这有助于在具有“开放表面上的射流”流动室的仪器中对前向和大角度光散射进行高灵敏度检测。前向散射在约2度以上的角度进行检测,而大角度散射包括18度以上的角度。理论上的考虑表明,为了根据细胞内结构区分细胞,在20度左右测量的大角度散射可能与在90度测量的一样有效。这一观点得到了以下发现的支持:使用基于弧光灯的仪器获得的白细胞悬浮液的双参数光散射直方图与使用大角度探测器位于90度的基于激光的仪器记录的此类直方图非常相似。不同种类的细菌可以通过双参数光散射装置进行区分,不同种类的海藻也可以。该装置的灵敏度足以在前向和大角度散射中测量0.2微米的聚苯乙烯颗粒。