Arch Orthop Trauma Surg. 2011 Dec;131(12):1731-7. doi: 10.1007/s00402-011-1360-z. Epub 2011 Aug 5.
Although wear reduction of highly cross-linked polyethylene has been demonstrated to be independent of femoral head diameter in some simulation studies, the clinical effects of femoral head diameter on wear of highly cross-linked polyethylene sockets remain unclear. We compared the in vivo wear of Longevity highly cross-linked polyethylene liners at 8-year follow-up using 26- or 32-mm cobalt-chromium heads.
A retrospective cohort study was performed on 90 cementless total hip arthroplasties (THAs) using Longevity highly cross-linked polyethylene liners combined with 26- (45 THAs) or 32-mm (45 THAs) cobalt-chromium heads. Annual radiographs were analyzed using PolyWare computer-assisted methods and linear and volumetric total head penetration rates and linear and volumetric steady-state wear rates were compared.
The two groups showed similar background data, and no significant differences were identified between groups in total head penetration rate or steady-state wear rate. Steady-state wear rates were negligible. Osteolysis was not observed in any hips in either group.
At the 8-year follow-up, wear of the Longevity was the same irrespective of the use 26- or 32-mm heads.
尽管一些模拟研究表明,高交联聚乙烯的磨损与股骨头直径无关,但股骨头直径对高交联聚乙烯窝的磨损的临床影响仍不清楚。我们比较了使用 26 或 32mm 钴铬头的 8 年随访时长寿高交联聚乙烯衬垫的体内磨损。
对 90 例使用长寿高交联聚乙烯衬垫联合 26mm(45 例)或 32mm(45 例)钴铬头的非骨水泥全髋关节置换术进行回顾性队列研究。使用 PolyWare 计算机辅助方法分析每年的射线照片,并比较总头穿透率的线性和体积以及稳态磨损率的线性和体积。
两组患者的背景数据相似,两组患者的总头穿透率或稳态磨损率无显著差异。稳态磨损率可以忽略不计。两组均未观察到任何髋关节的骨溶解。
在 8 年随访时,无论使用 26mm 或 32mm 头,长寿的磨损情况相同。