Invest Ophthalmol Vis Sci. 2011 Nov 17;52(12):8884-90. doi: 10.1167/iovs.10-6411.
To determine the interobserver and intraobserver measurement reproducibility of cornea parameters of both normal eyes and eyes with bullous keratopathy (BK) obtained with the Zhongshan Assessment Program (ZAP) on anterior segment optical coherence tomography (AS-OCT) images.
A comparative study was carried out on 24 healthy volunteers and 25 subjects with BK. AS-OCT images were independently analyzed by two examiners. Parameters examined: anterior chamber depth (ACD), central corneal thickness (CCT), posterior corneal curvature (PCC), and posterior corneal arc length (PCAL). Interobserver and intraobserver reproducibility of these parameters was calculated in terms of limits of agreement (mean of differences ± 1.96SD of differences).
In the normal group, both horizontal and vertical ACD were successfully measured in 23 of 24 (96%) images. The mean bias for two measurements by two different observers ranged from 0.003 to 0.117 mm for ACD, PCC, and PCAL measurements and from 0.013 to 2.25 μm for CCT measurements, and there were no differences between the two observers (P > 0.05). Mean bias for two measurements by the same grader ranged from 0.005 to 0.327 mm for ACD, PCC, and PCAL measurements and 1.46 to 2.53 μm for CCT measurements. There was no difference between the two observations (P > 0.05). Similar results were found in the BK group.
There was high inter- and intraobserver reproducibility for normal and pathologic corneas using the ZAP software. The ZAP software may serve as a new investigatory tool for accurately evaluating the anterior segment and corneal parameters for corneal procedures.
利用眼前节光学相干断层扫描(AS-OCT)图像中的中山评估程序(ZAP)确定正常眼和大疱性角膜病变(BK)眼中的角膜参数的观察者间和观察者内测量可重复性。
对 24 名健康志愿者和 25 名 BK 患者进行了一项比较研究。两名检查者独立分析了 AS-OCT 图像。检查的参数:前房深度(ACD)、中央角膜厚度(CCT)、后角膜曲率(PCC)和后角膜弧长(PCAL)。这些参数的观察者间和观察者内重复性通过差异的界限(差异的平均值±1.96SD 的差异)来计算。
在正常组中,24 个图像中的 23 个成功测量了水平和垂直 ACD。两个不同观察者的两次测量的平均偏差范围为 ACD、PCC 和 PCAL 测量值的 0.003 至 0.117 毫米,CCT 测量值的 0.013 至 2.25 微米,两个观察者之间没有差异(P > 0.05)。同一位评分员的两次测量的平均偏差范围为 ACD、PCC 和 PCAL 测量值的 0.005 至 0.327 毫米,CCT 测量值的 1.46 至 2.53 微米。两次观察之间没有差异(P > 0.05)。BK 组也得到了类似的结果。
使用 ZAP 软件,正常和病理角膜的观察者间和观察者内重复性均较高。ZAP 软件可作为评估眼前节和角膜参数的新研究工具,用于角膜手术。