Invest Ophthalmol Vis Sci. 2007 Aug;48(8):3683-8. doi: 10.1167/iovs.06-1120.

Abstract

PURPOSE

To evaluate the reproducibility of anterior chamber (AC) angle measurements obtained using anterior segment optical coherence tomography (AS-OCT).

METHODS

Patients with suspected glaucoma and those with glaucoma, ocular hypertension, or anatomically narrow angles were recruited from the glaucoma service at the National University Hospital, Singapore. All subjects underwent imaging of the nasal, temporal, and inferior AC angles with an AS-OCT prototype under standardized dark and light conditions. For short-term reproducibility analysis, a single observer acquired two sets of images followed by a third set of images acquired by a second observer. The interval between sessions was 10 minutes. For long-term reproducibility analysis, a single observer acquired two sets of images at least 24 hours apart. Images were measured using custom software to determine the AC depth (ACD), angle opening distance at 500 microm (AOD(500)), angle recess area at 500 microm (ARA(500)), and trabecular-iris space area at 500 microm (TISA(500)). The intraclass correlation coefficient (ICC) was calculated as a measure of intraobserver and interobserver reproducibility.

RESULTS

Twenty eyes of 20 patients were analyzed for short-term reproducibility, and 23 eyes of 23 patients were analyzed for long-term reproducibility. AC depth measurement demonstrated excellent reproducibility (ICC 0.93-1.00) in both dark and light conditions. For the nasal and temporal quadrants, all AC angle parameters demonstrated good to excellent short-term (ICC 0.67-0.90) and long-term (ICC 0.56-0.93) reproducibility in both dark and light conditions. In the inferior quadrant, reproducibility was lower in all categories of analysis and varied from poor to good (ICC 0.31-0.73).

CONCLUSIONS

AS-OCT allows quantitative assessment of the AC angle. The reproducibility of AC angle measurements was good to excellent for the nasal and temporal quadrants. The lower reproducibility of measurements in the inferior quadrant may be unique to this prototype due to difficulty in acquiring high-quality images of the inferior angle. Further assessment of the commercially available AS-OCT is needed to clarify this finding.

摘要

目的

评估使用眼前节光学相干断层扫描(AS-OCT)获得的前房角测量值的可重复性。

方法

从新加坡国立大学医院青光眼门诊招募疑似青光眼患者以及青光眼、高眼压症或解剖学上窄房角患者。所有受试者在标准化暗室和明室条件下,使用AS-OCT原型对鼻侧、颞侧和下方前房角进行成像。对于短期可重复性分析,由一名观察者采集两组图像,随后由另一名观察者采集第三组图像。两次检查之间的间隔为10分钟。对于长期可重复性分析,由一名观察者至少间隔24小时采集两组图像。使用定制软件测量图像,以确定前房深度(ACD)、500微米处的房角开放距离(AOD(500))、500微米处的房角隐窝面积(ARA(500))以及500微米处的小梁-虹膜间隙面积(TISA(500))。计算组内相关系数(ICC)作为观察者内和观察者间可重复性的指标。

结果

对20例患者的20只眼进行短期可重复性分析,对23例患者的23只眼进行长期可重复性分析。前房深度测量在暗室和明室条件下均显示出极好的可重复性(ICC 0.93 - 1.00)。对于鼻侧和颞侧象限,所有前房角参数在暗室和明室条件下均显示出良好至极好的短期(ICC 0.67 - 0.90)和长期(ICC 0.56 - 0.93)可重复性。在下方象限,所有分析类别的可重复性较低,从差到好不等(ICC 0.31 - 0.73)。

结论

AS-OCT可对前房角进行定量评估。鼻侧和颞侧象限的前房角测量可重复性良好至极好。下方象限测量的可重复性较低可能是该原型所特有的,原因是获取下方房角高质量图像存在困难。需要对市售AS-OCT进行进一步评估以阐明这一发现。