INTRODUCTION

The purpose of this study was to compare the effectiveness of debris removal between EndoVac, PiezoFlow, or needle irrigation (Max-i-Probe) in mandibular molars.

METHODS

The mesial roots of 30 extracted mandibular molars were mounted in resin by using the K-Kube and then sectioned at 2 and 4 mm from the apex. The specimens were reassembled and instrumented to size 40/0.04. A standard irrigation protocol was used during instrumentation in the PiezoFlow and Max-i-Probe groups. During instrumentation of teeth in the EndoVac group, the manufacturer's recommendations for irrigation were followed. Images of the canals and isthmuses were taken before final irrigation. The percentage of cleanliness for each canal and isthmus was calculated by using interactive software. The specimens were reassembled for final irrigation with EndoVac, PiezoFlow, and Max-i-Probe by using similar volumes of irrigants. Images were again used to calculate cleanliness. Comparisons of canal and isthmus cleanliness before and after final irrigation were made by using paired t tests, and the groups were compared with repeated-measures analysis of variance (P < .05).

RESULTS

There were no statistically significant differences in canal and isthmus cleanliness among all 3 groups at 2 and 4 mm from working length before or after final irrigation. Canal and isthmus cleanliness significantly improved at all levels after the final irrigation regimen in each group.

CONCLUSIONS

Final irrigation by using EndoVac, the PiezoFlow, or Max-i-Probe with similar volumes of irrigants significantly improved canal and isthmus cleanliness.

摘要

简介

本研究旨在比较 EndoVac、PiezoFlow 和针式冲洗(Max-i-Probe)在下颌磨牙中清除碎片的效果。

方法

将 30 颗离体下颌磨牙的近中根用 K-Kube 固定在树脂中,然后从根尖 2 和 4mm 处进行切割。将标本重新组装并锉至 40/0.04 号。在 PiezoFlow 和 Max-i-Probe 组中,在锉削过程中使用标准冲洗方案。在 EndoVac 组中,根据制造商的建议进行冲洗。在最后冲洗前拍摄根管和峡部的图像。通过交互式软件计算每个根管和峡部的清洁度百分比。用类似体积的冲洗剂重新组装最终冲洗的 EndoVac、PiezoFlow 和 Max-i-Probe。再次使用图像计算清洁度。使用配对 t 检验比较最终冲洗前后根管和峡部的清洁度,并使用重复测量方差分析(P<0.05)比较组间差异。

结果

在最终冲洗前后,2mm 和 4mm 工作长度处,所有 3 组的根管和峡部清洁度均无统计学差异。在每组的最终冲洗方案后,所有水平的根管和峡部清洁度均显著提高。

结论

使用类似体积的冲洗剂进行最终冲洗时,EndoVac、PiezoFlow 和 Max-i-Probe 均能显著提高根管和峡部的清洁度。