Jiang Guang-Ming, Wang Bao-Long, Wan Xiao-Ju, Wang Min, Zhou Jian-Hua, Liao Yan-Qiu
Department of Blood Transfusion, Anhui Provincial Hospital, Anhui Province, China.
Zhongguo Shi Yan Xue Ye Xue Za Zhi. 2011 Aug;19(4):1048-52.
The purpose of this study was to determine whether the fully automated ORTHO AutoVue Innova system, which based on the microcolumn glass sphere technology, is accurate enough to meet immunohematology testing needs at blood banks. 16 IgM anti-C, anti-c, anti-D, anti-E and anti-e dilution series were tested respectively, with corresponding antigen positive red blood cell solutions, by ORTHO AutoVue Innova system and saline medium test. 16 IgG anti-D dilution series were tested respectively with RhD positive red blood cell solutions by ORTHO AutoVue Innova system, polybrene test and antiglobulin test. The accuracies of microcolumn glass sphere technology were analysed, by comparing to the reference assays. The results showed that the sensitivities of the ORTHO AutoVue Innova tests were 1:69.8, 1:33.4, 1:1448.1, 1:139.6 and 1:32.0 for IgM anti-C, anti-c, anti-D, anti-E and anti-e respectively; the corresponding value of saline medium tests were 1:16.7, 1:16.6, 1:430.5, 1:34.9 and 1:9.9. There were statistically significant differences between the groups of each tests (t values were 14.38, 5.48, 10.25, 12.65 and 9.59 for IgM anti-C, anti-c, anti-D, anti-E and anti-e respectively, p < 0.05). For IgG anti-D, the sensitivities of the ORTHO AutoVue Innova test, polybrene test and antiglobulin test were 1:980.6, 1:181.0 and 1:304.4 respectively. There was statistically significant difference among the 3 groups (F = 51.15, p < 0.01). It is concluded the use of ORTHO AutoVue Innova system for blood group compatibility test can obtain more accurate results than traditional tube tests, it is reliable and safe for routine tests performed in immunohematology laboratories.
本研究的目的是确定基于微柱玻璃珠技术的全自动ORTHO AutoVue Innova系统是否足够准确,以满足血库免疫血液学检测需求。通过ORTHO AutoVue Innova系统和盐水介质试验,分别用相应的抗原阳性红细胞溶液对16个IgM抗 - C、抗 - c、抗 - D、抗 - E和抗 - e稀释系列进行检测。通过ORTHO AutoVue Innova系统、聚凝胺试验和抗球蛋白试验,分别用RhD阳性红细胞溶液对16个IgG抗 - D稀释系列进行检测。通过与参考检测方法比较,分析微柱玻璃珠技术的准确性。结果显示,对于IgM抗 - C、抗 - c、抗 - D、抗 - E和抗 - e,ORTHO AutoVue Innova试验的敏感性分别为1:69.8、1:33.4、1:1448.1、1:139.6和1:32.0;盐水介质试验的相应值分别为1:16.7、1:16.6、1:430.5、1:34.9和1:9.9。各试验组之间存在统计学显著差异(对于IgM抗 - C、抗 - c、抗 - D、抗 - E和抗 - e,t值分别为14.38、5.48、10.25、12.65和9.59,p < 0.05)。对于IgG抗 - D,ORTHO AutoVue Innova试验、聚凝胺试验和抗球蛋白试验的敏感性分别为1:980.6、1:181.0和1:304.4。三组之间存在统计学显著差异(F = 51.15,p < 0.01)。得出结论,使用ORTHO AutoVue Innova系统进行血型相容性检测比传统试管检测能获得更准确的结果,在免疫血液学实验室进行常规检测时可靠且安全。