Defence R & D Establishment, Jhansi Road, Gwalior 474002, India.
Talanta. 2011 Sep 30;85(4):1812-7. doi: 10.1016/j.talanta.2011.07.008. Epub 2011 Jul 14.
A novel piezoelectric (PZ) immunosensor for the direct detection of malarial Plasmodium falciparum histidine rich protein-2 (PfHRP-2) antigen was developed. The mixed self-assembled monolayers (SAMs) of thioctic acid and 1-dodecanethiol were formed on gold surface of quartz crystal. Cyclic voltammetry, electrochemical impedance spectroscopy and surface Raman spectroscopy techniques were used to characterize the mixed SAMs. The rabbit anti-PfHRP-2 antibodies were coupled on mixed SAM modified gold surface of quartz crystal via NHS/EDC activation method. The PZ immunosensor was applied to detect PfHRP-2 in the linear range of 15-60 ng/ml with a detection limit of 12 ng/ml. It was also found that even after 14 days of storage, 50% of the activity still remained. Clinical human serum samples were tested with this method, and the results were in agreement with those obtained from commercially available ICT kit (NOW(®) Malaria).
一种用于直接检测疟原虫恶性疟原虫富组氨酸蛋白-2(PfHRP-2)抗原的新型压电(PZ)免疫传感器被开发出来。金硫辛酸和 1-十二硫醇的混合自组装单层(SAM)在石英晶体的金表面形成。循环伏安法、电化学阻抗谱和表面拉曼光谱技术用于表征混合 SAM。通过 NHS/EDC 活化法将兔抗 PfHRP-2 抗体偶联到石英晶体混合 SAM 修饰的金表面上。该 PZ 免疫传感器用于检测 PfHRP-2,线性范围为 15-60ng/ml,检测限为 12ng/ml。还发现,即使在储存 14 天后,仍有 50%的活性保持。该方法用于检测临床人血清样本,结果与市售 ICT 试剂盒(NOW(®) Malaria)一致。