Suppr超能文献

利用超声体模建立肝癌筛查超声检查质量保证测试的截断值:韩国全国性调查的初步报告。

Establishing cutoff values for a quality assurance test using an ultrasound phantom in screening ultrasound examinations for hepatocellular carcinoma: an initial report of a nationwide survey in Korea.

机构信息

Department of Radiology, Seoul St Mary's Hospital, the Catholic University of Korea, Seoul, Korea.

出版信息

J Ultrasound Med. 2011 Sep;30(9):1221-9. doi: 10.7863/jum.2011.30.9.1221.

Abstract

OBJECTIVES

The purpose of this study was to evaluate the results of ultrasound (US) device testing using a US phantom and to determine cutoff values for phantom quality assurance tests in US examinations for the screening of hepatocellular carcinoma in Korea.

METHODS

Ultrasound phantom images were acquired from the general hospitals in Korea that participated in the National Cancer Screening Program for hepatocellular carcinoma. Ultrasound images of the phantom were acquired with a 3.0- to 5.0-MHz convex transducer and evaluated in terms of the dead zone, vertical and horizontal measurement, axial and lateral resolution, sensitivity, and gray scale/dynamic range. Appropriate cutoff values were determined to guarantee minimal qualifications for the performance of the US scanners.

RESULTS

Three hundred fifty-seven US scanners were tested using the following cutoff values: less than 2 mm for the dead zone, 5% discrepancy in the vertical measurement, 7.5% discrepancy in the horizontal measurement, all 11 identifiable line targets for axial and lateral resolution, more than 14 cm for sensitivity, and more than 4 cylindrical structures for gray scale/dynamic range. With these criteria, 283 US scanners (79.3%) passed the tests. The most common cause of disqualification was the dynamic range/gray scale. No statistical difference was observed in the disqualification rate between 3 groups based on different years of manufacture.

CONCLUSIONS

Through this study, we have defined cutoff values for phantom images acquired with US scanners. These will be used in performing screening US examinations for hepatocellular carcinoma in Korea.

摘要

目的

本研究旨在评估使用超声(US)体模进行 US 设备测试的结果,并确定韩国用于肝细胞癌筛查的 US 检查中体模质量保证测试的截止值。

方法

从参与国家癌症筛查计划的韩国综合医院获取超声体模图像。使用 3.0-5.0MHz 凸阵换能器获取体模的超声图像,并从盲区、垂直和水平测量、轴向和侧向分辨率、灵敏度和灰度/动态范围等方面进行评估。确定适当的截止值,以保证 US 扫描仪的性能达到最低要求。

结果

使用以下截止值对 357 台 US 扫描仪进行了测试:盲区小于 2mm、垂直测量相差 5%、水平测量相差 7.5%、轴向和侧向分辨率的所有 11 个可识别线靶、灵敏度大于 14cm、灰度/动态范围大于 4 个圆柱结构。符合这些标准的 283 台 US 扫描仪(79.3%)通过了测试。最常见的不合格原因是动态范围/灰度。根据不同的制造年份,三组之间的不合格率没有统计学差异。

结论

通过本研究,我们确定了使用 US 扫描仪获取的体模图像的截止值。这些将用于在韩国进行肝细胞癌筛查的超声检查。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验