累积和技术提供的胎儿颈项透明层测量的连续独立质量控制。

Continuous independent quality control for fetal nuchal translucency measurements provided by the cumulative summation technique.

机构信息

Frauenheilkunde, UniversitätsSpital Zürich, Zürich.

出版信息

Ultraschall Med. 2011 Dec;32 Suppl 2:E141-6. doi: 10.1055/s-0031-1281669. Epub 2011 Aug 29.

Abstract

PURPOSE

The cumulative summation technique (CUSUM) is an innovative method for the quality control of nuchal translucency (NT) measurements. CUSUM allows immediate corrective intervention as soon as an unacceptable tendency is noted. The aim of this study was to implement an objective and dynamic quality control method based on the CUSUM technique for prompt analysis of fetal NT measurement which would be compatible with different standards in routine clinical practice. The findings were compared to the standard NT quality control methods currently in use.

MATERIALS AND METHODS

Three sets of fetal NT measurements performed by three experienced examiners (I, II and III) were selected for retrospective evaluation. One additional set of NT measurements performed by examiner IV was prospectively assessed to approve the practicability of the method. NT measurements were conducted according to the recommendations of Fetal Medical Foundation (FMF) Germany and London. NT values were converted to Z-scores. For quality and accuracy evaluation, data were fed into the Digisono CUSUM software to create double CUSUM charts of Z-scores. In addition, histograms were composed from the Z-scores of each set of measurements and plotted against a normal Gaussian distribution.

RESULTS

Three different patterns of retrospective performance and one set of NT measurements that was evaluated prospectively are presented. The full alignment of Z-scores using CUSUM curves reflected exact periods of under- and overestimation of NT measurements. The CUSUM chart of the prospective data set reveals that prompt corrective intervention of poor performance resulted in reconstitution of optimal results and provided sufficient control. In contrast, histograms of NT Z-scores only showed a minor positive or negative shift as compared to the expected values on the basis of Gaussian distribution, but could not identify poor performance.

CONCLUSION

Use of the CUSUM technique analysing the quality of sonographic NT measurements provides the possibility to prospectively observe the development of the examiner's skills, to maintain competence and to promptly define the time when inaccurate measurements start to occur.

摘要

目的

累积和技术(CUSUM)是一种用于颈项透明层(NT)测量质量控制的创新方法。CUSUM 允许在发现不可接受的趋势时立即进行纠正干预。本研究的目的是基于 CUSUM 技术实施一种客观和动态的质量控制方法,以便及时分析胎儿 NT 测量值,使其与常规临床实践中的不同标准兼容。将研究结果与当前使用的标准 NT 质量控制方法进行比较。

材料和方法

选择了三位经验丰富的检查者(I、II 和 III)进行的三组胎儿 NT 测量值进行回顾性评估。第四位检查者进行的一组额外的 NT 测量值进行了前瞻性评估,以验证该方法的实用性。NT 测量按照德国胎儿医学基金会(FMF)和伦敦的建议进行。NT 值转换为 Z 分数。为了进行质量和准确性评估,将数据输入到 Digisono CUSUM 软件中,以创建 Z 分数的双 CUSUM 图表。此外,从每组测量值的 Z 分数组成直方图,并绘制在正态高斯分布上。

结果

呈现了三种不同的回顾性表现模式和一组前瞻性评估的 NT 测量值。CUSUM 曲线的 Z 分数完全对齐反映了 NT 测量值的低估和高估的确切时期。前瞻性数据集的 CUSUM 图表显示,对不良表现的及时纠正干预导致最佳结果的重建,并提供了充分的控制。相比之下,NT Z 分数的直方图仅显示与基于高斯分布的预期值相比的较小正或负偏移,但无法识别不良表现。

结论

使用 CUSUM 技术分析超声 NT 测量值的质量提供了前瞻性观察检查者技能发展的可能性,以保持能力并及时确定不准确测量开始发生的时间。