Nexfin无创逐搏动脉血压监测在自主神经功能测试中的应用。

Application of Nexfin noninvasive beat-to-beat arterial blood pressure monitoring in autonomic function testing.

作者信息

Sipkens Laura M, Treskes Kaij, Ariese-Beldman Karin, Veerman Derk P, Boer Christa

机构信息

Department of Anesthesiology, Institute for Cardiovascular Research, VU University Medical Center, Amsterdam, The Netherlands.

出版信息

Blood Press Monit. 2011 Oct;16(5):246-51. doi: 10.1097/MBP.0b013e32834b4431.

Abstract

OBJECTIVE

Evaluation of autonomic function responses is increasingly important for risk prediction and hemodynamic evaluation in the ambulant and perioperative setting, but requires a noninvasive arterial blood pressure measurement device. This study describes whether a novel noninvasive beat-to-beat arterial blood pressure measurement device (Nexfin HD) is able to reproducibly reflect autonomic function responses in healthy volunteers.

METHODS

Noninvasive beat-to-beat arterial blood pressure measurements (Nexfin HD) were performed in 20 healthy men of 22 ± 3 years. Measurements were performed during supine steady state, controlled breathing (0.125 Hz), passive leg raising, a controlled Valsalva maneuver, and a quick stand test. Finally, relative changes in pulse pressure during autonomic function testing and the test-retest reproducibility were determined.

RESULTS

Autonomic function tests induced beat-to-beat arterial blood pressure changes that were accurately monitored by the Nexfin device. The intraclass correlation coefficients for systolic and diastolic arterial blood pressure measurements during supine steady state were agreeable [0.91 (0.82-0.96) and 0.84 (0.69-0.93), respectively]. The reproducibility of blood pressure changes during controlled breathing, passive leg raising, and Valsalva maneuver averaged 0.92 (0.82-0.96), 0.76 (0.50-0.90), and 0.94 (0.89-0.97), respectively. The reproducibility of the pulse pressure variation (PPV) as calculated from controlled breathing-induced changes in the arterial blood pressure (13 ± 5%) was high [0.96 (0.93-0.98)].

CONCLUSION

This study shows that noninvasive beat-to-beat Nexfin HD arterial blood pressure measurements reproducibly reflect autonomic function responses in healthy volunteers.

摘要

目的

在门诊及围手术期环境中,评估自主神经功能反应对于风险预测和血流动力学评估愈发重要,但这需要一种无创动脉血压测量设备。本研究描述了一种新型的逐搏无创动脉血压测量设备(Nexfin HD)是否能够在健康志愿者中可重复地反映自主神经功能反应。

方法

对20名年龄在22±3岁的健康男性进行逐搏无创动脉血压测量(Nexfin HD)。测量在仰卧位稳态、控制呼吸(0.125赫兹)、被动抬腿、控制性瓦尔萨尔瓦动作和快速站立试验期间进行。最后,确定自主神经功能测试期间脉压的相对变化以及重测的可重复性。

结果

自主神经功能测试引起的逐搏动脉血压变化由Nexfin设备准确监测。仰卧位稳态期间收缩压和舒张压测量的组内相关系数令人满意[分别为0.91(0.82 - 0.96)和0.84(0.69 - 0.93)]。控制呼吸、被动抬腿和瓦尔萨尔瓦动作期间血压变化的可重复性平均分别为0.92(0.82 - 0.96)、0.76(0.50 - 0.90)和0.94(0.89 - 0.97)。由控制呼吸引起的动脉血压变化计算得出的脉压变异(PPV)的可重复性较高[0.96(0.93 - 0.98)]。

结论

本研究表明,无创逐搏Nexfin HD动脉血压测量能够在健康志愿者中可重复地反映自主神经功能反应。