Department of Radiodiagnosis, Sanjay Gandhi Post Graduate Institute of Medical Sciences, Lucknow, India.
AJNR Am J Neuroradiol. 2011 Dec;32(11):2120-5. doi: 10.3174/ajnr.A2674. Epub 2011 Sep 8.
DWI by using higher b-values provides tissue diffusivity with less T2 shinethrough effect. VCD in the abscess cavity correlates with ADC values. The purpose of this study was to investigate which b-value-derived ADC correlates better with VCD.
Thirty patients with brain abscess underwent conventional MR imaging and DWI with b = 1000, 2000, and 3000 s/mm(2) on a 3T MR imaging scanner. ADC values were quantified by placing regions of interest inside the abscess cavity in all sections where the lesion was apparent on coregistered ADC maps derived from different b-values. VCD was measured on pus aspirated.
An increase in b-value was associated with a decrease in ADC values in normal parenchyma as well as in the abscess cavity. The most significant negative correlation of VCD was observed with b = 3000 s/mm(2) (r = -0.98, P = .01).
VCD in the abscess cavity can be best assessed at b = 3000 s/mm(2) secondary to the reduction in the T2 shinethrough effect. DWI with b = 3000 s/mm(2) is of promising value in the assessment of the therapeutic response of brain abscess.
使用较高 b 值的 DWI 可提供 T2 穿透效应较小的组织扩散度。脓肿腔内的 VCD 与 ADC 值相关。本研究旨在探讨哪个 b 值衍生的 ADC 值与 VCD 的相关性更好。
30 例脑脓肿患者在 3T 磁共振成像扫描仪上进行常规磁共振成像和 b 值为 1000、2000 和 3000 s/mm²的 DWI。通过在源自不同 b 值的 ADC 图配准的所有病变明显的切片上,在脓肿腔内放置感兴趣区来量化 ADC 值。在抽吸脓液时测量 VCD。
b 值的增加与正常实质以及脓肿腔内的 ADC 值降低相关。与 VCD 相关性最强的是 b = 3000 s/mm²(r = -0.98,P =.01)。
由于 T2 穿透效应的降低,在 b = 3000 s/mm² 时可最佳评估脓肿腔内的 VCD。b = 3000 s/mm² 的 DWI 在评估脑脓肿的治疗反应方面具有很大的应用价值。