新型钛锆牙科种植体的组织整合:犬的比较组织学和放射学研究。

Tissue integration of a new titanium-zirconium dental implant: a comparative histologic and radiographic study in the canine.

机构信息

Clinic for Fixed and Removable Prosthodontics and Dental Material Science, University of Zurich, Zurich, Switzerland.

出版信息

J Periodontol. 2011 Oct;82(10):1453-61. doi: 10.1902/jop.2010.100737.

Abstract

BACKGROUND

This study evaluates a newly developed titanium-zirconium implant (TiZr), comparing it to a commercially available pure titanium (Ti) implant subjected to the same surface treatment.

METHODS

In nine dogs, 12 implants (six TiZr and six Ti) were randomly placed in the mandible with the implant shoulder at the bone crest and subjected to submerged healing. Standardized radiographs were taken after implantation, and at the sacrifice of 2 weeks (three dogs), 4 weeks (three dogs), and 8 weeks (three dogs). Histologic and histomorphometric measurements were performed on non-decalcified histologic sections. The main outcome measures included the first bone-implant contact (fBIC) and BIC over time. For statistical analysis, Wilcoxon signed-rank test and mixed model regressions were applied.

RESULTS

From baseline to 8 weeks, a mean bone loss of 0.09 ± 0.33 mm for TiZr and a gain of 0.02 ± 0.33 mm for Ti were calculated radiographically. The number of implants with the fBIC coronal to the reference point (implant shoulder) gradually increased over time, reaching 39% of all TiZr implants and 50% of all Ti implants at 8 weeks. The mean fBIC values for Ti and TiZr were 0.29 ± 0.42 mm and 0.26 ± 0.32 mm (2 weeks), -0.01 ± 0.20 mm and 0.10 ± 0.28 mm (4 weeks), and -0.06 ± 0.22 mm and 0.08 ± 0.30 mm (8 weeks), respectively. The mean BIC values peaked at 86.9% ± 6.8% (8 weeks) for TiZr and at 83.4% ± 5.9% (4 weeks) for Ti. No statistically significant differences were observed at any time point.

CONCLUSION

TiZr and Ti bone level implants with chemically-modified, sandblasted, and acid-etched surfaces performed similarly in regards to osseointegration in this unloaded canine study.

摘要

背景

本研究评估了一种新开发的钛锆(TiZr)种植体,并将其与经过相同表面处理的市售纯钛(Ti)种植体进行比较。

方法

在 9 只狗中,随机将 12 个种植体(TiZr 种植体 6 个,Ti 种植体 6 个)植入下颌骨,种植体肩部位于牙槽嵴顶,并进行埋置愈合。植入后、处死前 2 周(3 只狗)、4 周(3 只狗)和 8 周(3 只狗)分别拍摄标准 X 线片。对非脱钙组织学切片进行组织学和组织形态计量学测量。主要观察指标包括初始骨-种植体接触(fBIC)和随时间的 BIC。统计分析采用 Wilcoxon 符号秩检验和混合模型回归。

结果

从基线到 8 周,TiZr 的平均骨损失为 0.09±0.33mm,Ti 的平均骨获得为 0.02±0.33mm。在 X 线片上,种植体 fBIC 与参考点(种植体肩部)冠状位的种植体数量逐渐增加,8 周时,TiZr 种植体达到 39%,Ti 种植体达到 50%。Ti 和 TiZr 的平均 fBIC 值分别为 0.29±0.42mm 和 0.26±0.32mm(2 周)、-0.01±0.20mm 和 0.10±0.28mm(4 周)和-0.06±0.22mm 和 0.08±0.30mm(8 周)。TiZr 的平均 BIC 值在 8 周时达到 86.9%±6.8%,Ti 的平均 BIC 值在 4 周时达到 83.4%±5.9%。在任何时间点均未观察到统计学差异。

结论

在这项非负荷性犬研究中,具有化学改性、喷砂酸蚀表面的 TiZr 和 Ti 骨水平种植体在骨整合方面表现相似。