Department of Mechanical Engineering, Biomechanical Laboratory, Las Palmas University, Las Palmas de Gran Canaria, Spain.
Med Eng Phys. 2012 Jul;34(6):717-24. doi: 10.1016/j.medengphy.2011.09.015. Epub 2011 Oct 7.
This paper analyses whether it is possible to use dynamic compression plates (DCPs) and screw locking elements (SLEs) to vary the flexibility of osteoporotic fracture fixation without compromising the strength and stability of the construct. Compression, torsion and four-point bending static strength tests were conducted. Cyclic load tests of up to 10,000 load cycles were also carried out to determine stiffness performance. Four fixation systems were mounted onto polyurethane bone models. Group 1 consists of the DCP and six cortical screws. Group 2, idem, but with the addition of two SLEs. Group 3, idem, but with the addition of six SLEs. Group 4 used the locking compression plate (LCP) and locking screws. The results indicated no significant difference (p>0.05) in the strength of groups 2-4. It was also observed that the torsional stiffness of group 3 (0.30 Nm/°) was higher than that of group 2 (0.23 Nm/°) and similar to that of group 4 (0.28 Nm/°). Compression stiffness of group 4 (124 N/mm) was higher than that of group 2 (102 N/mm), but lower than that of group 3 (150 N/mm). No notable differences were observed for structural bending stiffness. It is concluded that by using the DCP with SLEs it is possible to modify the stiffness of the fixation construct for the repair of osteoporotic fractures and, in this way, facilitate the conditions suitable on secondary bone healing.
本文分析了在不影响结构强度和稳定性的情况下,是否可以使用动力加压钢板(DCP)和螺钉锁定元件(SLE)来改变骨质疏松性骨折固定的灵活性。进行了压缩、扭转和四点弯曲静态强度测试。还进行了多达 10000 次循环负载测试,以确定其刚度性能。将四个固定系统安装在聚氨酯骨模型上。第 1 组由 DCP 和六根皮质螺钉组成。第 2 组与第 1 组相同,但增加了两个 SLE。第 3 组与第 1 组相同,但增加了六个 SLE。第 4 组使用锁定加压钢板(LCP)和锁定螺钉。结果表明,第 2-4 组的强度没有显著差异(p>0.05)。还观察到,第 3 组(0.30 Nm/°)的扭转刚度高于第 2 组(0.23 Nm/°),与第 4 组(0.28 Nm/°)相似。第 4 组(124 N/mm)的压缩刚度高于第 2 组(102 N/mm),但低于第 3 组(150 N/mm)。结构弯曲刚度没有明显差异。结论是,使用带 SLE 的 DCP 可以改变骨质疏松性骨折固定的刚度,从而为二次骨愈合创造有利条件。