Chiriac S, Facca S, Diaconu M, Gouzou S, Liverneaux P
Department of Hand Surgery, University Hospital of Strasbourg, 10 avenue Baumann, Illkirch cedex, France.
J Hand Surg Eur Vol. 2012 May;37(4):342-9. doi: 10.1177/1753193411422685. Epub 2011 Oct 10.
Synthetic nerve guides are occasionally used to repair nerve defects. The aim of the present work was to analyse the results of Neurolac™ use in a series of 23 patients. We operated on 28 nerve lesions located on various sites: arm (n = 1), elbow (n = 5), forearm (n = 4), wrist (n = 2), palm (n = 5), fingers (n = 11). Defects averaged 11.03 mm and were repaired using Neurolac™. After an average of 21.9 months' follow up (3-45 months), subjective criteria (pain, cold intolerance, Quick DASH) and objective criteria (strength, Weber and Semmes-Weinstein sensitivity tests) were compared with the contralateral side. Average pain score was 2.17/10. Cold intolerance was reported in fifteen cases. Quick DASH averaged 35.37/100. Grip strength averaged 64.62% of the contralateral side. As regards sensitivity, the difference between the two sides was 18.89 on Weber's test, and 46.92 on Semmes-Weinstein. Defect size did not affect the outcomes. We observed eight complications the most serious being two fistulizations of the Neurolac™ device close to a joint and one neuroma. Neurolac™ presents some advantages (resorption, semi-permeability, emergency use, tenseless repair) like other synthetic guides used for nerve regeneration and its transparency constitutes an added benefit. However, some difficulty in its handling and its expensiveness represent real disadvantages. Our results are not in favour of its use in repairing hand nerve defects.
合成神经导管偶尔用于修复神经缺损。本研究的目的是分析在一系列23例患者中使用Neurolac™的结果。我们对位于不同部位的28处神经损伤进行了手术:手臂(n = 1)、肘部(n = 5)、前臂(n = 4)、手腕(n = 2)、手掌(n = 5)、手指(n = 11)。缺损平均为11.03毫米,使用Neurolac™进行修复。平均随访21.9个月(3 - 45个月)后,将主观标准(疼痛、冷不耐受、Quick DASH)和客观标准(力量、Weber和Semmes - Weinstein感觉测试)与对侧进行比较。平均疼痛评分为2.17/10。15例报告有冷不耐受。Quick DASH平均为35.37/100。握力平均为对侧的64.62%。在感觉方面,两侧在Weber测试中的差异为18.89,在Semmes - Weinstein测试中为46.92。缺损大小不影响结果。我们观察到8例并发症,最严重的是Neurolac™装置在关节附近出现2例瘘管形成和1例神经瘤。Neurolac™具有一些优点(吸收、半透性、紧急使用、无张力修复),与其他用于神经再生的合成导管一样,其透明度是一个额外优势。然而,其操作存在一些困难且价格昂贵,这是实际的缺点。我们的结果不支持将其用于修复手部神经缺损。