Suppr超能文献

硬脑膜缝合的密闭性研究:不同密封剂的体外比较。

Study of dural suture watertightness: an in vitro comparison of different sealants.

机构信息

Department of Neurosurgery, Groupe Hospitalier Pitié-Salpêtrière and Lariboisiere Hospital, 29 avenue Jean Moulin, 75014 Paris, France.

出版信息

Acta Neurochir (Wien). 2011 Dec;153(12):2465-72. doi: 10.1007/s00701-011-1197-9. Epub 2011 Oct 12.

Abstract

BACKGROUND

CSF leakages constitute a major complication of intradural procedures, especially for posterior fossa and skull base surgery. Dural suture watertightness is a decisive issue, and neurosurgeons routinely use different products to reinforce their dural closure. We have designed an experimental system capable of testing CSF leak pressure levels in order to compare two types of sutures in vitro and particularly four different sealants.

METHODS

Twenty-five fresh human cadaveric dural samples were removed and prepared for testing in a pressure chamber system connected to a hydraulic pressure motor. CSF leak levels were objectively registered. First, simple interrupted stitches were compared to running simple closure on 50-mm linear suture. Secondly, four sealants (two sealants/glues, Bioglue®, Duraseal®; two haemostatics, Tachosil®, Tissucol®) were tested. Statistical analysis was performed with paired Student's t-test.

RESULTS

No significant difference between interrupted closure and running suture was observed (p = 0.079). All sealants increased the watertightness of the suture significatively. However, comparison of the means of the differences for each product revealed large variations. In the conditions of our experiment, one sealant (Duraseal®) and one haemostatic (Tachosil®) seemed to show better results. We observed two different types of leakage: at the dura-sealant interface and through the sealant itself.

CONCLUSIONS

We have developed an experimental device capable of testing dural closure watertightness. Interrupted stitch suturing seemed no different from running simple closure. On the contrary, the sealants tested show different watertightness capacities.

摘要

背景

CSF 漏出是硬脊膜内手术的主要并发症,尤其是在后颅窝和颅底手术中。硬脑膜缝合的密闭性是一个决定性的问题,神经外科医生通常使用不同的产品来加强硬脑膜的闭合。我们设计了一个实验系统,能够测试 CSF 泄漏压力水平,以便在体外比较两种类型的缝线,特别是四种不同的密封剂。

方法

从 25 具新鲜的人体硬脑膜标本中取出并准备在连接到液压马达的压力室系统中进行测试。CSF 泄漏水平被客观地记录下来。首先,比较了简单间断缝合与 50mm 线性缝合的连续缝合。其次,测试了四种密封剂(两种密封胶/胶,Bioglue ® ,Duraseal ® ;两种止血剂,Tachosil ® ,Tissucol ® )。采用配对学生 t 检验进行统计学分析。

结果

间断缝合与连续缝合之间无显著性差异(p=0.079)。所有的密封剂都显著增加了缝线的密闭性。然而,对每种产品的差异平均值进行比较显示出很大的差异。在我们实验的条件下,一种密封剂(Duraseal ® )和一种止血剂(Tachosil ® )似乎显示出更好的效果。我们观察到两种不同类型的泄漏:在硬脑膜-密封剂界面和通过密封剂本身。

结论

我们开发了一种能够测试硬脑膜闭合密闭性的实验装置。间断缝合似乎与连续简单缝合没有区别。相反,测试的密封剂显示出不同的密闭能力。