Radiology, NorthShore University HealthSystem, Evanston, Illinois 60201, USA.
J Magn Reson Imaging. 2012 Feb;35(2):309-17. doi: 10.1002/jmri.22835. Epub 2011 Oct 11.
To enhance the reliability and spatial resolution of magnetization transfer ratio (MTR) measurements for interrogation of subcortical brain regions with an automated volume of interest (VOI) approach.
A 3D magnetization transfer (MT) sequence was acquired using a scan-rescan imaging protocol in nine healthy volunteers. VOI definition masks for the MTR measurements were generated using FreeSurfer and compared to a manual region of interest (ROI) approach. (The longitudinal stability of MTR was monitored using agar gel phantom over a 5-month period.) Intraclass correlation coefficients (ICCs), coefficients of variation (CVs), and instrumental standard deviation (ISD) were determined.
CVs ranged from 1.29%-2.64% (automated) vs. 1.30%-3.40% (manual). ISDs ranged from 0.62-1.10 pu (automated) vs. 0.68-1.67 pu (manual). The SD of the running difference was 1.70% for the phantom scans. The Bland-Altman method indicated interchangeability of the automated VOI and manual ROI measurements.
The automated VOI approach for MTR measurement yielded higher ICCs, lower CVs, and lower ISDs compared to the manual method, supporting the utility of this strategy. These results demonstrate the feasibility of obtaining reliable MTR measurements in hippocampus and other critical subcortical regions.
通过自动化感兴趣区(VOI)方法,提高磁共振磁化转移率(MTR)测量在皮质下脑区检测中的可靠性和空间分辨率。
在 9 名健康志愿者中使用扫描-扫描成像方案采集三维磁化转移(MT)序列。使用 FreeSurfer 生成 MTR 测量的 VOI 定义掩模,并与手动 ROI 方法进行比较。(使用琼脂凝胶体在 5 个月的时间内监测 MTR 的纵向稳定性。)确定了组内相关系数(ICC)、变异系数(CV)和仪器标准偏差(ISD)。
CV 范围为 1.29%-2.64%(自动)与 1.30%-3.40%(手动)。ISD 范围为 0.62-1.10 pu(自动)与 0.68-1.67 pu(手动)。对于体模扫描,运行差异的 SD 为 1.70%。Bland-Altman 方法表明,自动 VOI 和手动 ROI 测量具有可互换性。
与手动方法相比,MTR 测量的自动 VOI 方法产生了更高的 ICC、更低的 CV 和更低的 ISD,支持该策略的实用性。这些结果表明,在海马体和其他关键皮质下区域获得可靠的 MTR 测量是可行的。