According to the Italian laws, some categories of workers entrusted with duties possibly constituting a threat to security, physical safety, and health of third parties have to be screened to exclude the use/abuse of the following drugs of abuse: opiates, cocaine, cannabinoids, amphetamine, methamphetamine, 3,4-methylenedioxymethamphetamine, methadone, and buprenorphine. Toxicological tests can be performed with urinary on-site rapid screening devices, provided that sensitivities up to specified cutoffs are ensured. The present study reports performances, in terms of sensitivity, specificity, and accuracy, of an automatic on-site test and of an FPIA-based device, using gas chromatography/mass spectrometry (GC/MS) as a reference methodology. Three levels of concentration were tested, corresponding to the cutoff and to 2 and 3 times the limits, respectively. In terms of sensitivities, neither the on-site nor the benchtop instrumentations gave positive results, since values of zero percentage were obtained for concentrations up to 2-fold the limits. Even if good results were obtained in terms of specificity and accuracy by both devices, none of them seem to be adequate for the current application to the toxicological screening at workplaces. In fact, a rapid screening device can be used for drug tests provided that it ensures sensitivity at the prescribed cutoffs. Data showed that such is completely rejected and a more sensitive instrumentation should be preferred.
根据意大利法律,一些可能对第三方的安全、身体安全和健康构成威胁的职责类别的工人必须接受筛查,以排除使用/滥用以下滥用药物:阿片类药物、可卡因、大麻素、苯丙胺、甲基苯丙胺、3,4-亚甲二氧基甲基苯丙胺、美沙酮和丁丙诺啡。毒理学测试可以使用尿液现场快速筛选设备进行,前提是确保达到指定截止值的灵敏度。本研究报告了自动现场测试和基于 FPIA 的设备的性能,以气相色谱/质谱 (GC/MS) 为参考方法,在灵敏度、特异性和准确性方面。测试了三个浓度水平,分别对应于截止值和 2 倍和 3 倍的限制。就灵敏度而言,现场和台式仪器都没有给出阳性结果,因为在 2 倍限制以下的浓度下,得到的都是零百分比的值。尽管两种设备在特异性和准确性方面都取得了良好的结果,但它们都似乎不适合当前在工作场所进行毒理学筛查的应用。事实上,快速筛选设备可用于药物测试,前提是它在规定的截止值处确保灵敏度。数据表明,这种方法完全被拒绝,应该优先选择更敏感的仪器。