The potential for mouthguards to change the risk of concussion was studied in football helmet impacts. The Hybrid III head was modified with an articulating mandible, dentition, and compliant temporomandibular joints (TMJ). It was instrumented for triaxial head acceleration and triaxial force at the TMJs and upper dentition. Mandible force and displacement were validated against cadaver impacts to the chin. In phase 1, one of five mouthguards significantly lowered HIC in 6.7 m/s impacts (p = 0.025) from the no mouthguard condition but not in 9.5 m/s tests. In phase 2, eight mouthguards increased HIC from +1 to +17% in facemask impacts that loaded the chinstraps and mandible; one was statistically higher (p = 0.018). Peak head acceleration was +1 to +15% higher with six mouthguards and 2-3% lower with two others. The differences were not statistically significant. Five of eight mouthguards significantly reduced forces on the upper dentition by 40.8-63.9%. Mouthguards tested in this study with the Hybrid III articulating mandible lowered forces on the dentition and TMJ, but generally did not influence HIC or concussion risks.
本文研究了防护牙套对口颌部冲击所致脑震荡风险的改变作用。Hybrid III 头模配备了可活动的下颌、牙齿和顺应性颞下颌关节(TMJ)。头模的三轴加速度和 TMJ 及上牙列的三轴力都经过了测量。下颌力和位移通过与尸体对下巴的冲击进行了验证。在第一阶段,在 6.7 m/s 冲击(无防护牙套条件下 p = 0.025)中,五种防护牙套中有一种显著降低了 HIC,但在 9.5 m/s 测试中则没有。在第二阶段,八种防护牙套增加了面罩冲击时加载下颌带和下颌的 HIC(+1 至+17%);其中一种统计学上更高(p = 0.018)。六个防护牙套的峰值头部加速度增加了+1 至+15%,而另外两个则降低了 2-3%。这些差异没有统计学意义。八种防护牙套中有五种可显著降低上牙列受力 40.8-63.9%。本研究中使用 Hybrid III 可活动下颌的八种防护牙套降低了牙齿和 TMJ 的受力,但通常不会影响 HIC 或脑震荡风险。