to evaluate the accuracy of cone beam computed tomography (CBCT) for volumetric measurement of teeth, using micro-computed tomography (Micro-CT) as the reference standard.
The Department of Orthodontics at Sichuan University. The sample consisted of 27 maxillary and mandibular premolars of 15 patients, planned to be extracted for orthodontic treatment.
The 27 teeth were subjected to standardized CBCT scanning before extraction and Micro-CT scanning after extraction. From CBCT data, teeth were tissue segmented and then three-dimensionally (3D) reconstructed, while from Micro-CT data, teeth were 3D reconstructed directly. Tooth volumes were then calculated. The intra-observer repeatability and reproducibility of two observers and the overall between-instrument agreement of the measurements were evaluated using intra-class correlation coefficients (ICCs) and concordance correlation coefficient (CCC), respectively.
The intra-observer repeatability was high for both observers. ICCs were 0.999 and 0.998, respectively. The reproducibility of the two observers was also high (ICC, 0.740). The overall between-instrument agreement of the measurements was good, and CCC was 0.993 and its lower 95% confidence interval was 0.989.
The accuracy of the CBCT method for volumetric measurement of teeth in vivo is comparable to the Micro-CT method in vitro. The CBCT method has the potential possibility to be applied in studies on root resorption associated with orthodontic force. Further study is needed to prove the sensitivity of the method.
以 micro-computed tomography (Micro-CT) 为参考标准,评估锥形束 CT (CBCT) 用于牙齿体积测量的准确性。
四川大学正畸科。样本由 15 名患者的 27 颗上颌和下颌前磨牙组成,计划因正畸治疗而拔除。
在拔牙前对 27 颗牙齿进行标准化的 CBCT 扫描,在拔牙后进行 Micro-CT 扫描。从 CBCT 数据中,对牙齿进行组织分割,然后进行三维(3D)重建,而从 Micro-CT 数据中,直接对牙齿进行 3D 重建。然后计算牙齿体积。使用组内相关系数(ICC)和一致性相关系数(CCC)分别评估两位观察者的组内重复性和可重复性以及测量的总体仪器间一致性。
两位观察者的组内重复性均较高。ICC 分别为 0.999 和 0.998。两位观察者的可重复性也较高(ICC,0.740)。测量的总体仪器间一致性较好,CCC 为 0.993,其 95%置信区间下限为 0.989。
CBCT 方法在体内测量牙齿体积的准确性与体外 Micro-CT 方法相当。CBCT 方法有可能应用于研究正畸力相关的牙根吸收。需要进一步的研究来证明该方法的敏感性。