Liu Yi, Mah James, Xu Tian-min
Department of Orthodontics, Peking University School and Hospital of Stomatology, Beijing 100081, China.
Beijing Da Xue Xue Bao Yi Xue Ban. 2010 Feb 18;42(1):98-102.
To determine the accuracy of volumetric analysis of teeth in vivo using cone beam computed tomography (CBCT).
The physical volume (Vw) of 24 bicuspids extracted for orthodontic purposes (16 were imaged with the i-CAT and 8 with the CB MercuRay before extraction) was determined using water displacement technique. Corresponding pre-treatment CBCT image data were uploaded into Amira 4.0 for segmentation and radiographic volume (Va). All measurements were performed twice by two observers. The statistical difference between Vw and Va was assessed using a paired t-test. The intra- and inter-observer reliability was determined by calculating Pearson correlation coefficients and intraclass correlation coefficients (ICC).
The overall mean Vw of teeth specimens was (0.553 + or - 0.082) cm(3), while the mean Va was (0.548 + or - 0.079) cm(3), (0.529 + or - 0.078) cm(3) for observer 1, and (0.567 + or - 0.085) cm(3) for observer 2. There were statistically significant differences between Va and Vw (P<0.05). Between two observers, Va had statistically significant difference (P<0.05). The inter- and intra-observers correlation coefficient for Vw was high. Lastly, surface smoothing reduced the volume by 3%-12%.
In vivo determination of tooth volumes from CBCT data is feasible. The measurements slightly deviate from the physical volumes within -4% to 7%. Smoothing operations reduce volume measurements.
使用锥形束计算机断层扫描(CBCT)确定体内牙齿体积分析的准确性。
采用排水法测定24颗因正畸目的拔除的双尖牙的物理体积(Vw)(其中16颗在拔除前用i-CAT成像,8颗用CB MercuRay成像)。将相应的治疗前CBCT图像数据上传至Amira 4.0进行分割并获得影像学体积(Va)。两名观察者对所有测量均进行两次。使用配对t检验评估Vw和Va之间的统计学差异。通过计算Pearson相关系数和组内相关系数(ICC)来确定观察者内和观察者间的可靠性。
牙齿标本的总体平均Vw为(0.553±0.082)cm³,而平均Va为(0.548±0.079)cm³,观察者1测得的平均Va为(0.529±0.078)cm³,观察者2测得的平均Va为(0.567±0.085)cm³。Va和Vw之间存在统计学显著差异(P<0.05)。两名观察者之间,Va存在统计学显著差异(P<0.05)。Vw的观察者间和观察者内相关系数较高。最后,表面平滑使体积减少了3%-12%。
从CBCT数据体内确定牙齿体积是可行的。测量值与物理体积略有偏差,偏差在-4%至7%之间。平滑操作会减少体积测量值。