Johnson Casey Jl, Christianson Chad D, Sheaff Chrystal N, Laine Derek F, Zimmer Jennifer Sd, Needham Shane R
Alturas Analytics, Inc., 1324 Alturas Drive, Moscow, ID 83843, USA.
Bioanalysis. 2011 Oct;3(20):2303-10. doi: 10.4155/bio.11.212.
Conventional liquid-handling devices were employed, along with an improved punching device, to semi-automate dried blood spot (DBS) extraction of alprazolam, α-hydroxyalprazolam and midazolam from human whole blood. Liquid-handling devices were used to add internal standard to the DBS cards and to extract the analytes from the DBS, in order to be analyzed by HPLC-MS/MS.
The technique was shown to be accurate (±12.0%) and precise (10.3%) across the dynamic range of the assay.
The semi-automated extraction reduced sample preparation time by more than 50% when compared with more conventional DBS manual extraction methods.