Musculoskeletal Quantitative Imaging Research Group, Department of Radiology and Biomedical Imaging, University of California, San Francisco, San Francisco, CA, USA.
Bone. 2012 Jan;50(1):111-8. doi: 10.1016/j.bone.2011.10.003. Epub 2011 Oct 13.
Motion artifacts are a common finding during high-resolution peripheral quantitative computed tomography (HR-pQCT) image acquisitions. To date it is not clear (i) when to repeat an acquisition, (ii) when to exclude a motion-degraded dataset post hoc, and (iii) how motion induced artifacts impact measures of trabecular and cortical parameters. In this study we present inter- and intra-observer reproducibility of a qualitative image quality grading score and report the prevalence of repeat acquisitions in our population. Finally the errors in bone density and micro-architectural parameters estimated from repeat acquisitions with and without motion degradation are presented. The relationship between these errors and the image quality grade is evaluated for each parameter. Repeat acquisitions performed due to operator-observed motion in the reconstructed image occurred for 22.7% of the exams (29.7% radius, 15.7% tibia). Of this subset, 88 exams with repeat acquisitions had at least one acquisition graded 1 (best quality). In this subset, the percent differences in bone density and micro-architecture measures tended to increase as the relative image quality decreased. Micro-architectural parameters were more sensitive to motion compared to geometric and densitometric parameters. These results provide estimates of the error in bone quality measures due to motion artifacts and provide an initial framework for developing standardized quality control criteria for cross-sectional and longitudinal HR-pQCT studies.
运动伪影是高分辨率外周定量 CT(HR-pQCT)图像采集的常见现象。迄今为止,尚不清楚(i)何时需要重复采集,(ii)何时需要事后排除运动退化的数据集,以及(iii)运动引起的伪影如何影响骨小梁和皮质参数的测量。在这项研究中,我们展示了定性图像质量分级评分的观察者内和观察者间可重复性,并报告了我们人群中重复采集的发生率。最后,报告了存在和不存在运动退化的重复采集的骨密度和微结构参数的估计误差。对于每个参数,都评估了这些误差与图像质量等级之间的关系。由于重建图像中操作员观察到的运动,有 22.7%的检查(桡骨 29.7%,胫骨 15.7%)需要重复采集。在这部分中,有 88 次重复采集的检查至少有一次评分达到 1(最佳质量)。在这个子集里,骨密度和微结构测量的百分比差异往往随着相对图像质量的降低而增加。与几何和密度参数相比,微结构参数对运动更敏感。这些结果提供了由于运动伪影导致的骨质量测量误差的估计,并为制定用于横断面和纵向 HR-pQCT 研究的标准化质量控制标准提供了初步框架。