Second Radiology Clinic İzmir Katip Çelebi University, Atatürk Training and Research Hospital, İzmir, Turkey.
Diagn Interv Radiol. 2012 Mar-Apr;18(2):147-52. doi: 10.4261/1305-3825.DIR.4526-11.1. Epub 2011 Sep 24.
To evaluate, in terms of their protective features, the lead aprons used in areas working with ionizing radiation at a hospital by analyzing qualitative and quantitative aspects using a variety of methods.
Eighty-five protective lead aprons used in our hospital's clinics to work with ionizing radiation were analyzed in the radiology unit. Each apron was identified by registering the unit from which it had been obtained and by how long it had been used, its storage condition, and its lead thickness. X-ray films of the aprons, controlled according to their appearances, durability and cleanliness, were taken to evaluate their internal structure; their permeability was measured with electronic dosimeters in terms of their absorbent features. All of these data were compared with the results acquired from brand-new, Turkish Standards Institution approved aprons having different lead thicknesses.
Regarding internal structure homogeneity, only 13 (15.3%) of 85 aprons were found to be at normal levels and usable. A total of 14 (16.5%) of the remaining 72 aprons' radiation absorptions were at normal levels, but folds were observed in their protective lead layers. The remaining 58 aprons (68.2%) were found to be defective. All of the aprons were considered to be defective in terms of their radiation permeability.
All of the aprons were found insufficient for protection and were more radioparent than the defined limits; it was concluded that they must be replaced by new ones.
通过使用多种方法分析定性和定量方面,评估医院放射科使用的电离辐射防护铅衣的防护性能。
在放射科分析了我院临床科室使用的 85 件防护铅衣。每件铅衣都通过登记其来源单位以及使用时间、储存条件和铅厚度进行了识别。对外观、耐用性和清洁度进行了控制的铅衣 X 射线片,以评估其内部结构;通过电子剂量计测量其渗透性,以评估其吸收特性。将所有这些数据与具有不同铅厚度的全新、符合土耳其标准协会的铅衣的结果进行了比较。
仅 13 件(15.3%)的 85 件铅衣被认为是正常水平且可使用的。在其余 72 件铅衣中,共有 14 件(16.5%)的辐射吸收率正常,但防护铅层出现了褶皱。其余 58 件(68.2%)被认为有缺陷。所有的铅衣在辐射渗透性方面都被认为是有缺陷的。
所有的铅衣都被认为是防护不足的,比规定的限制更具放射性;因此,必须用新的铅衣来替换。